Elberling C
OTICON Research Centre, Eriksholm, Snekkersten, Denmark.
J Am Acad Audiol. 1999 May;10(5):248-60.
The present work was undertaken in an attempt to evaluate whether it is reasonable to expect that categorical loudness scaling can provide useful information for nonlinear hearing aid fitting. Normative data from seven scaling procedures show that the individual procedures relate the perceptual categories differently to sound level and with a substantial between-subject variance. Hearing-impaired data from four studies demonstrate that the inverse slope of the loudness function varies linearly with hearing loss and with a constant variance. In relation to hearing aid fitting, the slope can, in most cases, be predicted from the hearing loss with an accuracy within the range of a normal finetuning. For the fitting of nonlinear hearing aids, the statistical properties of both normal and impaired loudness functions are equally important. The present analysis strongly suggests that categorical loudness scaling cannot, in general, provide significant information for the fitting process.
开展本研究是为了评估期望分类响度标度能为非线性助听器验配提供有用信息是否合理。来自七种标度程序的规范数据表明,各个程序将感知类别与声级的关联方式不同,且个体间差异很大。来自四项研究的听力受损者数据表明,响度函数的反斜率随听力损失呈线性变化,且方差恒定。对于助听器验配而言,在大多数情况下,可根据听力损失预测斜率,其准确度在正常微调范围内。对于非线性助听器的验配,正常和受损响度函数的统计特性同样重要。目前的分析强烈表明,一般而言,分类响度标度无法为验配过程提供重要信息。