使用通道清洁刷处理功能异常的Tenckhoff导管。

The use of a channel-cleaning brush for malfunctioning Tenckhoff catheters.

作者信息

Kumwenda M J, Wright F K

机构信息

Glan Clwyd Hospital, Bodelwyddan, Clwyd, UK.

出版信息

Nephrol Dial Transplant. 1999 May;14(5):1254-7. doi: 10.1093/ndt/14.5.1254.

Abstract

BACKGROUND

Tenckhoff catheter malfunction causes morbidity in some patients on continuous ambulatory peritoneal dialysis (CAPD). Various methods of treatment of malfunction have been described; we report our experience of the use of a channel-cleaning brush for this problem.

METHOD

Ten patients on CAPD who developed catheter malfunction were identified. The causes of malfunction were catheter migration in one patient, catheter obstruction due to fibrin and clots in six, probable omental wrapping in one, and pain on draining of PD fluid in two patients. A channel-cleaning brush was manipulated repeatedly in and out of the catheters using aseptic technique and fluoroscopy guidance in an attempt to restore patency or dislodge the catheter to another site.

RESULTS

CAPD was successfully re-established after this procedure in eight patients (80%), two catheters were removed from those patients in whom the use of the brush was unsuccessful.

CONCLUSION

In this study a channel-cleaning brush was effective and safe in the treatment of Tenckhoff catheter malfunction.

摘要

背景

在一些持续非卧床腹膜透析(CAPD)患者中,Tenckhoff导管故障会导致发病。已经描述了各种处理故障的方法;我们报告了使用通道清洁刷解决此问题的经验。

方法

确定了10例发生导管故障的CAPD患者。故障原因包括1例导管移位,6例因纤维蛋白和血凝块导致的导管阻塞,1例可能是网膜包裹,以及2例患者在引流腹膜透析液时出现疼痛。在无菌技术和荧光透视引导下,将通道清洁刷反复插入和拔出导管,试图恢复通畅或使导管移位到另一个位置。

结果

8例患者(80%)在此操作后成功重新建立了CAPD,2例使用刷子不成功的患者的导管被拔除。

结论

在本研究中,通道清洁刷在治疗Tenckhoff导管故障方面有效且安全。