Suppr超能文献

Capillary isoelectric focusing with whole column imaging detection for analysis of proteins and peptides.

作者信息

Mao Q, Pawliszyn J

机构信息

Department of Chemistry, University of Waterloo, Ontario, Canada.

出版信息

J Biochem Biophys Methods. 1999 Feb 25;39(1-2):93-110. doi: 10.1016/s0165-022x(99)00006-8.

Abstract

Whole column imaging detection has been developed for capillary isoelectric focusing (CIEF) of proteins and peptides. In this imaged C1EF technique, a solution of sample and ampholytes was introduced into a short (4-5 cm), internally coated capillary stabilized by a cartridge. After applying high DC voltage, the isoelectric focusing process takes place and the focused zones are monitored in a real-time mode using the imaging detectors developed. Three types of imaging detectors have been developed including refractive index gradient, laser-induced fluorescence (LIF), and absorption. Of these, absorption imaging detection is the most practical at the present time due to its quantitative ability and universal characteristics. Whole column imaging detection eliminates the mobilization step required for single point detection after the focusing process. Therefore, it provides a fast analysis speed (3-5 min for each sample), and avoids the disadvantages associated with the mobilization process, such as distortion of pH gradient and loss in resolution. In this paper, we review the methodology of imaged CIEF as well as progress in instrumental development, IEF performed on a microchip, and the application to protein and peptide analysis.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验