Hilfiker P R, Quick H H, Debatin J F
Institute of Diagnostic Radiology, University Hospital Zurich, Switzerland.
Radiology. 1999 Jun;211(3):693-7. doi: 10.1148/radiology.211.3.r99jn23693.
To evaluate the characteristics of various plain and covered stents as regards demonstration of the contained lumen with simulated contrast material-enhanced fast three-dimensional (3D) magnetic resonance (MR) angiography.
Six stents (Easy Wallstent, Cragg, Palmaz, Cragg EndoPro System 1, Corvita, Passager) were implanted in plastic tubes and integrated into a closed-tubing circuit driven by a pulsatile roller flow pump. The circulating water was spiked with gadopentetate dimeglumine. Three-dimensional MR angiograms were obtained with an echo time of 1.4 or 2.1 msec and with partial (0.5 signal acquired) or full (one signal acquired) k-space sampling. The size of the stent-associated artifact was determined relative to the tube diameter.
The Easy Wallstent and Palmaz stent caused complete obliteration of the stent lumen with all sequences. The Cragg, Cragg EndoPro System 1, and Passager stents allowed good visualization of the stent lumen. The Corvita stent demonstrated major artifacts. The magnitude of the stent-associated artifact was related to the echo time (P < .01) but not to the type of k-space sampling (P = .35).
The luminal patency of selected plain and covered stents can be assessed with contrast-enhanced 3D MR angiography.
通过模拟对比剂增强快速三维(3D)磁共振(MR)血管造影术,评估各种普通支架和覆膜支架在显示其内部管腔方面的特点。
将六个支架(易威壁式支架、克拉格支架、帕尔马兹支架、克拉格腔内系统1型支架、科尔维塔支架、帕萨热支架)植入塑料管中,并集成到由脉动滚轴流泵驱动的封闭管路系统中。循环水中加入钆喷酸葡胺。采用回波时间为1.4或2.1毫秒,以及部分(采集0.5个信号)或完全(采集一个信号)k空间采样的方式获取三维MR血管造影图像。根据相对于管腔直径的大小来确定与支架相关的伪影大小。
在所有序列中,易威壁式支架和帕尔马兹支架导致支架管腔完全闭塞。克拉格支架、克拉格腔内系统1型支架和帕萨热支架能较好地显示支架管腔。科尔维塔支架显示出较大伪影。与支架相关的伪影大小与回波时间有关(P <.01),但与k空间采样类型无关(P =.35)。
通过对比剂增强3D MR血管造影术可评估选定的普通支架和覆膜支架的管腔通畅情况。