Guittet C, Ossant F, Vaillant L, Berson M
INSERM U316, faculté de médecine, Tours, France.
IEEE Trans Biomed Eng. 1999 Jun;46(6):740-6. doi: 10.1109/10.764950.
The aim of this study is in vivo skin tissue characterization of young and old human cutaneous tissues by estimating the slope of the attenuation coefficient. The method used is the centroid algorithm with a second-order autoregressive model to perform the spectral analysis. Backscattered signals are acquired with a 40-MHz transducer fixed on a three-dimensional robot. Diffraction phenomena are eliminated via an axial translation of the transducer that allows the acquisition of the signal in the focal zone. The slope of the attenuation coefficient is estimated on phantoms of known attenuation, in order to validate the method. Preliminary measurements of the slope of the attenuation coefficient are subsequently performed in the echographic mode on abdominal human skin samples in vitro at 40 MHz. After assessing the reproducibility of the measurement of the attenuation coefficient slope in human dermis at 40-MHz in vivo, this is carried out on the volar face of the forearm of 150 healthy subjects aged 14-85 yr. The values measured range from 0.7 to 3.6 dB/cm.MHz. The main result of this study is the decrease with advancing age of the attenuation coefficient slope, which may reflect structural modifications of human dermis with age.
本研究的目的是通过估计衰减系数的斜率,对年轻和老年人类皮肤组织进行体内皮肤组织特征分析。所采用的方法是使用二阶自回归模型的质心算法来进行频谱分析。使用固定在三维机器人上的40MHz换能器采集背向散射信号。通过换能器的轴向平移消除衍射现象,从而能够在聚焦区采集信号。在已知衰减的体模上估计衰减系数的斜率,以验证该方法。随后在40MHz体外超声模式下对人体腹部皮肤样本进行衰减系数斜率的初步测量。在评估了40MHz体内人体真皮中衰减系数斜率测量的可重复性后,对150名年龄在14 - 85岁的健康受试者的前臂掌面进行了测量。测量值范围为0.7至3.6dB/cm·MHz。本研究的主要结果是衰减系数斜率随年龄增长而降低,这可能反映了人类真皮随年龄的结构变化。