[验证一款7通道动态多导记录仪。1:医生和患者操作说明]

[Validating a 7-channel ambulatory polygraphy unit. 1: Operating instructions for the physician and patient].

作者信息

Verse T, Pirsig W, Kroker B, Junge-Hülsing B, Zimmermann E

机构信息

Sektion für Rhinologie und Rhonchopathien, Universitäts-HNO-Klinik Ulm.

出版信息

HNO. 1999 Apr;47(4):249-55.

Abstract

PURPOSE

For some time, the ambulatory diagnosis of sleep-related breathing disorders has included the use of seven-channel recording units. One such unit is the POLY-MESAM (MAP, Martinsried, FRG). The first part of the present study prospectively investigated the handling of the system for physicians and patients, its technical reliability, reliability of the software used and the results in comparison to handscoring.

METHODS

In all, 104 patients (95 males, 9 females) were studied for different severities of obstructive sleep-related breathing disorders. The first 51 patients were connected to the POLY-MESAM in the evening within the clinic. The patients then slept at home and returned the next day. Another 53 patients received only a short briefing in the clinic and connected themselves at home to the POLY-MESAM System. Each patient's status was monitored by means of a visual analogue scale. Automatic evaluation of the data was compared with the results of manual scoring.

RESULTS

From the data recorded only 6% of the results were not usable. Patients acceptances of the system were very high (94.3%). On average each patient required 21 minutes for attaching and detaching the device. Scoring and instructing the patients required an average of 23 min. Regarding the apnea-hypopnea index (AHI), the correlation between automatic (AHI = 11.1) and manual (AHI = 11.3) evaluations was high. Analysis of snoring turned out to be insufficient.

CONCLUSIONS

The POLY-MESAM proved to be reliable and user-friendly. Our findings show that the system can be recommended for outpatient screening of sleep-related breathing disorders, with patients able to manage the system without help. Validation by using simultaneous polysomnography is the subject of part two of the study.

摘要

目的

一段时间以来,睡眠相关呼吸障碍的门诊诊断一直采用七通道记录设备。其中一种设备是POLY - MESAM(MAP,德国马丁斯里德)。本研究的第一部分前瞻性地调查了该系统对于医生和患者的操作便利性、技术可靠性、所使用软件的可靠性以及与人工评分相比的结果。

方法

总共对104例患者(95例男性,9例女性)进行了不同严重程度的阻塞性睡眠相关呼吸障碍研究。前51例患者于晚上在诊所连接到POLY - MESAM。患者随后在家中睡眠,并于次日返回。另外53例患者仅在诊所接受简短介绍,然后在家中自行连接到POLY - MESAM系统。通过视觉模拟量表监测每位患者的状况。将数据的自动评估结果与人工评分结果进行比较。

结果

从记录的数据来看,只有6%的结果不可用。患者对该系统的接受度非常高(94.3%)。平均每位患者连接和拆卸设备需要21分钟。对患者进行评分和指导平均需要23分钟。关于呼吸暂停低通气指数(AHI),自动评估(AHI = 11.1)与人工评估(AHI = 11.3)之间的相关性很高。对打鼾的分析结果并不理想。

结论

POLY - MESAM被证明是可靠且用户友好的。我们的研究结果表明,该系统可推荐用于睡眠相关呼吸障碍的门诊筛查,患者无需帮助即可自行操作该系统。使用同步多导睡眠图进行验证是本研究第二部分的主题。