Suppr超能文献

主动脉瘤修复装置。

Devices for aortic aneurysm repair.

作者信息

May J, White G H, Harris J P

机构信息

Department of Surgery, University of Sydney, Australia.

出版信息

Surg Clin North Am. 1999 Jun;79(3):507-27. doi: 10.1016/s0039-6109(05)70021-9.

Abstract

Although the technical success and short-term efficacy of endovascular grafts have now been demonstrated, the long-term durability of grafts used in the aorta remains to be proven. No long-term data are yet available regarding device durability or patient outcome beyond the initial few years. Recent evidence shows that the ability of endovascular grafts to cause shrinkage and regression of aortic aneurysms may have a paradoxic effect of distorting the endograft itself, thus causing geometric changes within the supporting metallic framework and, ultimately, device failure (Fig. 9). Thus, the desired positive effect of the device may, ironically, lead to its eventual failure.

摘要

尽管血管内移植物的技术成功率和短期疗效现已得到证实,但用于主动脉的移植物的长期耐用性仍有待证明。目前尚无关于最初几年后设备耐用性或患者预后的长期数据。最近的证据表明,血管内移植物导致主动脉瘤收缩和缩小的能力可能会产生反常效应,使血管内移植物本身变形,从而在支撑金属框架内引起几何形状变化,并最终导致设备故障(图9)。因此,具有讽刺意味的是,该设备预期的积极效果可能会导致其最终失效。