Hanson W F, Grant W, Kennedy P, Cundiff J H, Gagnon W F, Berkley L W, Shalek R J
Med Phys. 1978 Nov-Dec;5(6):552-5. doi: 10.1118/1.594449.
One of the principal concerns of a physicist responsible for calibrating megavoltage radiotherapy equipment is the validity and stability of the 60Co exposure correction factor assigned to his ionization-chamber and electrometer system. It is the practice of the AAPM Radiological Physics Center (RPC) to perform an intercomparison between the RPC chamber and electrometer system and the chamber and electrometer in use at each of the various institutions visited by the RPC. The results of 202 such intercomparisons are reviewed to determine (1) the consistency in the assignment of exposure correction factors by a calibrating agency with itself and with other calibrating agencies, and (2) the dependence of the reliability of the exposure correction factors upon the type of field instrument and the time since calibration.