Müller H P, Gödde P, Czerski K, Agrawal R, Feilcke G, Reither K, Wolf K J, Oeff M
Department of Cardiology, University Clinic Benjamin Franklin, Free University Berlin, Germany.
J Med Eng Technol. 1999 May-Jun;23(3):108-15. doi: 10.1080/030919099294258.
The objective of this case report is to determine the accurate localization of a malignant ventricular tachycardia (VT) focus by combining multichannel magnetocardiographic (MCG) information with morphologic data. The localization was obtained by calculating the three-dimensional current density distribution (3D-CDD) on the left ventricular surface. To estimate the accuracy of this localization technique, examinations of a healthy volunteer were additionally performed. The MCG-signals were recorded in a magnetically shielded room by a 49-channel magnetogradiometer. The corresponding morphologic information was recorded by magnetic resonance tomography (MRT). The coordinate systems were matched with the help of markers. The 3D-CDD was calculated by the Philips CURRY software package. The origin of a malignant VT determined by X-ray images of the ablation catheter position during the electrophysiological examination (EPE), was used as the gold standard. This was then compared with the localization results obtained by the 3D-CDD. It was found that the localization coordinates showed a difference of less than 10 mm.