In continuation of our previously published report on initial experience with titanium implants in the middle ear [13], we now present the results of routine use over a three year period.
From November 1994 to November 1997, 661 titanium implants (Düsseldorf model) were implanted in the authors' clinics (355 PORP, 306 TORP). There was no preselection of patients. Five hundred eighty-two follow up examinations were performed with a medium follow up time of 11.6 months.
The good initial results of the previous study with respect to biocompatibility and functional hearing results were confirmed in the long term follow-up. The average air-bone gap was calculated over the frequencies 0.5, 1, 2, and 4 kHz. Closure to within 20 dB was achieved in 72% of cases. The air-bone gap tended to decrease with increasing time of implantation. Adverse reaction to the prostheses did not occur. Extrusion occurred in one case of complete middle ear atelectasis with resorption of the interposed cartilage. Insufficient improvement of hearing was attributable to a short implant in 12 cases (1.8%). In three cases (0.5%) insufficient stability of the PORP was attributable to eroded stapes suprastructure, and in three others (0.5%) a dislocation was responsible for a poor hearing result. Middle ear fibrosis with impairment of sound transmission was seen in 3 patients (0.5%).
The superior acoustic properties of the delicate yet rigid low-weight titanium implants combined with excellent biocompatibility lead to a good hearing result if a meticulous surgical technique is employed. The easy handling makes it a pleasure to work with these protheses.
继我们之前发表的关于中耳钛植入物初步经验的报告[13]之后,我们现在展示三年常规使用的结果。
从1994年11月至1997年11月,在作者的诊所植入了661个钛植入物(杜塞尔多夫模型)(355个部分听骨赝复物,306个全听骨赝复物)。未对患者进行预先筛选。进行了582次随访检查,平均随访时间为11.6个月。
先前研究在生物相容性和功能性听力结果方面的良好初步结果在长期随访中得到证实。在0.5、1、2和4千赫频率上计算平均气骨导差。72%的病例气骨导差缩小至20分贝以内。气骨导差倾向于随着植入时间的增加而减小。未发生对假体的不良反应。在一例完全性中耳肺不张伴插入软骨吸收的病例中发生了假体脱出。12例(1.8%)听力改善不足归因于植入物过短。3例(0.5%)部分听骨赝复物稳定性不足归因于镫骨上部结构侵蚀,另外3例(0.5%)听力结果不佳是由于脱位。3例患者(0.5%)出现中耳纤维化伴声音传导受损。
精致但坚固的低重量钛植入物具有卓越的声学特性,结合出色的生物相容性,如果采用细致的手术技术,可带来良好的听力结果。易于操作使得使用这些假体成为一种乐趣。