Suppr超能文献

检测 Björk-Shiley 凸凹型心脏瓣膜出口支柱单腿分离:使用两种不同的电影血管造影成像技术的共识图像采集与解读

Detection of Björk-Shiley convexo-concave heart valve outlet strut single leg separations: consensus image acquisition and interpretation using two different cineradiographic imaging technologies.

作者信息

Hopper K D, Gilchrist I C, Landis J R, Localio A R, Wilson R P, Pae W E, Kunselman A R, Griffith J W, Pierce W S, Potok P S, TenHave T R

机构信息

Department of Radiology, Department of Medicine, Penn State University, Hershey, Penn. 17033, USA.

出版信息

Cardiology. 1999;91(2):96-101. doi: 10.1159/000006887.

Abstract

BACKGROUND

Conventional cineradiology has been used clinically to detect partially broken outlet struts (single leg separations, SLSs) in normally functioning Björk-Shiley convexo-concave (C/C) heart valves. The value of radiographic screening has, however, been debated in the medical literature. This study uses the same radiographic technique in sheep implanted with known-status C/C valves in combination with a newly developed geometric image magnification radiography system. This study was designed to test whether sensitivity and specificity of radiographic screening of C/C valves in detecting SLSs could be improved through the combination of readers and imaging modalities.

METHODS

Twenty-one sheep with mitral C/C valves were studied on both systems. Five were used for extensive scanning training. When operators were expert on both systems, 16 blinded study valves (4 intact and 12 with outlet strut SLSs) were scanned twice on both systems, first on a modified conventional and then a prototype geometric image magnification (Feinfocus(TM)) cineradiographic system by two expert physicians working together.

RESULTS

Among the 32 scanned valves, the two combined expert physicians were required to evaluate 40 intact legs and 24 with an SLS. For all SLS valves, the conventional and Feinfocus systems separately detected 50 (12/24) and 54% (13/24), respectively. When the two systems were combined, the final consensus score was correct in 67% (16/24) of all SLS valves.

CONCLUSIONS

Combined modality, paired expert physicians detected 67% of all SLSs. The Feinfocus system might be best reserved for those patients in whom the Siemens screening study demonstrates in minimally suspicious (grade 2) or suspicious (grade 3) appearance of a C/C valve outlet strut leg.

摘要

背景

传统的电影放射成像技术已在临床上用于检测功能正常的 Björk-Shiley 凸凹型(C/C)心脏瓣膜中部分断裂的出口支柱(单腿分离,SLS)。然而,放射学筛查的价值在医学文献中一直存在争议。本研究在植入已知状态 C/C 瓣膜的绵羊中使用相同的放射学技术,并结合新开发的几何图像放大放射成像系统。本研究旨在测试通过读者和成像方式的组合,C/C 瓣膜放射学筛查检测 SLS 的敏感性和特异性是否可以提高。

方法

在两个系统上对 21 只植入二尖瓣 C/C 瓣膜的绵羊进行研究。5 只用于广泛的扫描训练。当操作员对两个系统都熟练掌握后,16 个盲法研究瓣膜(4 个完整的和 12 个有出口支柱 SLS 的)在两个系统上各扫描两次,首先在改良的传统系统上,然后在原型几何图像放大(Feinfocus(TM))电影放射成像系统上,由两位专家医生共同进行。

结果

在 32 个扫描的瓣膜中,两位联合专家医生需要评估 40 个完整的支柱和 24 个有 SLS 的支柱。对于所有有 SLS 的瓣膜,传统系统和 Feinfocus 系统分别检测出 50%(12/24)和 54%(13/24)。当两个系统结合时,最终的共识评分在所有有 SLS 的瓣膜中 67%(16/24)是正确的。

结论

联合模式下,配对的专家医生检测出了所有 SLS 的 67%。Feinfocus 系统可能最适合那些西门子筛查研究显示 C/C 瓣膜出口支柱腿有最低限度可疑(2 级)或可疑(3 级)外观的患者。