The standardization of 62Cu and experimental determinations of dose calibrator settings for generator-produced 62CuPTSM.

作者信息

Zimmerman B E, Cessna J T

机构信息

Physics Laboratory, National Institute of Standards and Technology, Gaithersburg, MD 20899, USA.

出版信息

Appl Radiat Isot. 1999 Nov;51(5):515-26. doi: 10.1016/s0969-8043(99)00078-0.

Abstract

The Positron Emission Tomography agent 62Cu Pyruvaldehyde Bis(N4-methyl)thiosemicarbazone (PTSM) has been standardized by the National Institute of Standards and Technology (NIST) using 4 pi beta liquid scintillation counting with 3H-standard efficiency tracing. Using a measurement model developed at NIST for the determination of experimental dose calibrator dial settings for short-lived radionuclides, the correct dial settings for 62CuPTSM in a clinically useful geometry (a 35 ml plastic syringe containing 33 ml of solution) as well as the standard NIST 5 ml glass ampoule have been elucidated. This measurement model is fully described, as is the treatment of associated uncertainties. The correct instrument settings for the NIST Capintec CRC-12 re-entrant ionization chamber ('dose calibrator') were found to be '499 +/- 6' and '489 +/- 8', respectively, for the syringe and ampoule geometries. The stated uncertainties are expanded (k = 2) uncertainties. The expanded uncertainties in the measured activity arising from these new dial settings are +/- 0.8% and +/- 1.4%, respectively, for the syringe and ampoule. The measured activities using these settings are lower than those obtained from the manufacture's recommended setting of '448' by 9.7% for the syringe and 7.7% for the ampoule, and underscore the need for experimental verification of the dose calibrator settings for each radionuclide and each geometry used in the clinic.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验