Suppr超能文献

自膨式覆膜支架移植物变形使血管腔内周围动脉瘤修复术复杂化。

Deformation of self-expanding stent-grafts complicating endovascular peripheral aneurysm repair.

作者信息

Sitsen M E, Ho G H, Blankensteijn J D

机构信息

Department of Vascular Surgery, University Hospital Utrecht, The Netherlands.

出版信息

J Endovasc Surg. 1999 Aug;6(3):288-92. doi: 10.1583/1074-6218(1999)006<0288:DOSSCE>2.0.CO;2.

Abstract

PURPOSE

To demonstrate the deformation of self-expandable stents after endovascular repair of peripheral aneurysms.

METHODS AND RESULTS

The Corvita Endoluminal Graft was used to treat a traumatic false aneurysm of the right subclavian artery and a common iliac artery aneurysm in 2 patients. In the subclavian case, the stent-graft showed a "cigar-shaped" deformation with hemodynamically significant stenoses at the proximal and distal ends at 3 months. In the second case, the same type of deformity was noted only 1 day after implantation. Two months later, the stent-graft occluded, necessitating surgical repair.

CONCLUSIONS

Both cases demonstrate the possibility of stent deformation of self-expanding stent-grafts implanted at arterial sites not subject to external compression.

摘要

目的

展示周围动脉瘤血管内修复术后自膨式支架的变形情况。

方法与结果

使用Corvita腔内移植物治疗2例患者,1例为右锁骨下动脉创伤性假性动脉瘤,另1例为髂总动脉瘤。在锁骨下动脉病例中,支架移植物在3个月时呈“雪茄形”变形,近端和远端出现血流动力学上有显著意义的狭窄。在第二例中,植入后仅1天就发现了相同类型的畸形。两个月后,支架移植物闭塞,需要进行手术修复。

结论

两例均表明在未受外部压迫的动脉部位植入自膨式支架移植物时,支架有可能发生变形。