Study on causes and prevention of electrostatic charge build-up during extracorporeal circulation.

作者信息

Snijders J, de Bruijn P, Bergmans M, Bastianen G

机构信息

Department of Clinical Perfusion, Ignatius Hospital, Breda, The Netherlands.

出版信息

Perfusion. 1999 Sep;14(5):363-70. doi: 10.1177/026765919901400508.

Abstract

For many years it has been known that roller pumps interfere regularly with the electrocardiogram (ECG) signal. This electrical activity is also considered responsible for the electrostatic charge build-up in some oxygenators with a plastic heat exchanger. During use of such an oxygenator, an electrical shock can be felt when touching the device. In the absence of a provoked discharge, a spontaneous discharge may occur. The construction of the heat exchanger and the use of nonconductive materials enable some devices to function as a capacitor. If the voltage difference across the fibre wall is too high, an electrostatic discharge may occur. In 1997, four blood-to-water leaks in Maxima Forté oxygenators were reported on a worldwide basis. After thorough research by Medtronic Cardiac Surgery (Cardiopulmonary Division), the cause of the leaks was found to be an electrothermal event across the fibre wall of one single fibre. No patient injuries were reported in any of the events. Medtronic patented a number of US Food and Drug Administration-approved and patient-safe solutions to prevent this electrostatic charge build-up. We studied the electrostatic phenomenon in order to clarify the clinical aspects. The in vitro part of the study is related to the behaviour of the Medtronic Biopump and the Stockert Shiley roller pump--in combination with the use of a PVC or silicone pumpheader. Subsequently, we have investigated the influence of the rotation or flow of both types of blood pumps and occlusion settings on the roller pump. Furthermore, we tried to capture the electrical charge generated by the roller pumps in a capacitor and in some oxygenators with different types of heat exchanger. Finally, we tested the external CEL (Charge Equalization Line) and the internal 'Gold Wire' charge equalization systems, both patented by Medtronic. In the in vivo part of the study, 15 patients were divided into three groups of five patients each. In the first group a roller pump with a PVC pumpheader was evaluated and, in the second group, a roller pump with a silicone pumpheader. The third group consisted of the Medtronic Biopump. In all the groups, a Maxima Forté oxygenator with the external CEL protection was used. Owing to the fact that, at the time, the Maxima Forté 'Gold Wire' oxygenator was not available for clinical use, only the external CEL could be evaluated.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验