Suppr超能文献

Detection of lens proteins in phaco tips after phacoemulsification.

作者信息

Nuyts R M, Van Boekel M, Bloemendal H, Hiddema F

机构信息

Department of Ophthalmology, University Eye Clinic, Maastricht, The Netherlands.

出版信息

J Cataract Refract Surg. 1999 Nov;25(11):1510-4. doi: 10.1016/s0886-3350(99)00241-2.

Abstract

PURPOSE

To determine the efficacy of cleaning procedures to remove water-soluble and water-insoluble lens proteins from phaco tips after phacoemulsification.

SETTING

Academic Hospital Maastricht, Department of Ophthalmology, University Eye Clinic, and Rotterdam Hospital, Rotterdam, The Netherlands.

METHODS

Detection of alpha A-crystallins, cytoskeletal lens proteins (vimentin), and lens-cell membranes (MP26) was performed on new and reused phaco tips with specific markers. To detect antibodies, sodium dodecyl sulfate-polyacrylamide gel electrophoresis and Western blotting techniques were used.

RESULTS

Threshold detection levels for alpha A-crystallins, vimentin, and MP26 were 100, 43, and 108 ng, respectively. No detectable residues of water-soluble or water-insoluble lens proteins were found in outflow samples of routinely cleaned phaco tips or in outflow samples of phaco tips that were not cleaned after phacoemulsification. Urea effectively eliminated water-insoluble lens proteins from contaminated phaco tips.

CONCLUSION

After regular cleaning of phaco tips, no detectable lens protein residues were found in the outflow samples. Moreover, omitting the cleaning of phaco tips after phacoemulsification did not lead to detectable lens proteins in the outflow samples.