Suppr超能文献

药物对手持式血糖仪和便携式血糖分析仪测量血糖的影响。

Effects of drugs on glucose measurements with handheld glucose meters and a portable glucose analyzer.

作者信息

Tang Z, Du X, Louie R F, Kost G J

机构信息

Department of Medical Pathology, School of Medicine, University of California, Davis 95616, USA.

出版信息

Am J Clin Pathol. 2000 Jan;113(1):75-86. doi: 10.1309/QAW1-X5XW-BVRQ-5LKQ.

Abstract

Thirty drugs used primarily in critical care and hospital settings were tested in vitro to observe interference on glucose measurements with 6 hand-held glucose meters and a portable glucose analyzer. Paired differences of glucose measurements between drug-spiked samples and unspiked control samples were calculated to determine bias. A criterion of +/- 6 mg/dL was used as the cutoff for interference. Ascorbic acid interfered with the measurements on all glucose devices evaluated. Acetaminophen, dopamine, and mannitol interfered with glucose measurements on some devices. Dose-response relationships help assessment of drug interference in clinical use. High dosages of these drugs may be given to critically ill patients or self-administered by patients without medical supervision. Package inserts for the glucose devices may not provide adequate warning information. Hence, we recommend that clinicians choose glucose devices carefully and interpret results cautiously when glucose measurements are performed during or after drug interventions.

摘要

对主要用于重症监护和医院环境的30种药物进行了体外测试,以观察其对6款手持式血糖仪和1台便携式葡萄糖分析仪测量葡萄糖的干扰情况。计算加药样品与未加药对照样品之间葡萄糖测量值的配对差异,以确定偏差。以±6mg/dL作为干扰的临界值。维生素C干扰了所有评估的葡萄糖检测设备的测量。对乙酰氨基酚、多巴胺和甘露醇在某些设备上干扰了葡萄糖测量。剂量反应关系有助于评估临床使用中的药物干扰。这些药物的高剂量可能会用于重症患者,或由患者在无医疗监督的情况下自行服用。葡萄糖检测设备的包装说明书可能未提供足够的警示信息。因此,我们建议临床医生在药物干预期间或之后进行葡萄糖测量时,谨慎选择葡萄糖检测设备并谨慎解读结果。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验