Mueller O, Hahnenberger K, Dittmann M, Yee H, Dubrow R, Nagle R, Ilsley D
Agilent Technologies, Waldbronn, Germany.
Electrophoresis. 2000 Jan;21(1):128-34. doi: 10.1002/(SICI)1522-2683(20000101)21:1<128::AID-ELPS128>3.0.CO;2-M.
Microfabrication technology was used to develop a system consisting of disposable glass chips containing etched channels, reagents including polymer matrix and size standards, computer-controlled instrumentation for performing electrophoretic separations and fluorescence detection of double-stranded DNA, and software for automated data analysis. System performance was validated for separation and quantitation reproducibility using samples varying in amount and size of DNA fragments, buffer composition, and salt concentrations. Several applications of the microfluidic system for DNA analysis have been demonstrated, such as of polymerase chain reaction (PCR) products, sizing of plasmid digests, and detection of point mutations by restriction fragment length polymorphism (RFLP) mapping.