Intermittent atrial level right-to-left shunt with temporary hypoxemia in a patient during support with a left ventricular assist device.

作者信息

Kilger E, Strom C, Frey L, Felbinger T W, Pichler B, Tichy M, Rank N, Wheeldon D, Kesel K, Schmitz C, Reichenspurner H, Polasek J, Weis F, Goetz A E

机构信息

Department of Anesthesiology, University of Munich, Germany.

出版信息

Acta Anaesthesiol Scand. 2000 Jan;44(1):125-7. doi: 10.1034/j.1399-6576.2000.440122.x.

Abstract

We report a 56-year-old male patient developing hypoxemia after surgical replacement of infected valves of a left ventricular assist device (LVAD, Novacor) which had supported him during the previous 15 months. Contrast transesophageal echocardiography (TEE) revealed an atrial septal defect with intermittent right-to-left shunt across a patent foramen ovale. We postulate that the shunt detected in this patient occurred as a consequence of reduced pulmonary vascular compliance due to positive end-expiratory pressure (PEEP) and an increase of mean intrathoracic pressure. Furthermore, we hypothesize that synchronized LVAD operation exacerbates any potential right-to-left shunt due to the profound left ventricular unloading which occurs during LVAD support. In this first report of a right-to-left shunt from a previously unrecognized patent foramen ovale in a Novacor patient, the subsequent transient hypoxemia could be managed by avoiding PEEP of more than 3 mmHg, and mean airway pressure of more than 11 mmHg and by careful volume replacement in order to prevent the pump from completely emptying the left ventricle (LV) and the left atrium (LA). Thus, prior to every LVAD implantation a transesophageal contrast echocardiography with Valsalva maneuver should be performed to identify intracardiac right-to-left shunt.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验