Shapiro G C, Leibowitz D W, Oz M C, Weslow R G, Di Tullio M R, Homma S
Division of Cardiology, Columbia University, New York, N.Y., USA.
J Heart Lung Transplant. 1995 May-Jun;14(3):594-7.
Intracardiac right-to-left shunting through a patent foramen ovale is a known cause of arterial hypoxemia. We present a case report of a patient supported with a left ventricular assist device who had significant right-to-left shunting as visualized with transesophageal echocardiography. When the device was turned off, no further shunting occurred and arterial hypoxemia resolved. Our report is the first visual representation of the anatomy of a patent foramen ovale in a patient supported with a left ventricular assist device.
经未闭卵圆孔的心脏内右向左分流是动脉血氧不足的一个已知原因。我们报告一例使用左心室辅助装置的患者,经食管超声心动图显示存在显著的右向左分流。当装置关闭时,分流不再发生,动脉血氧不足得到缓解。我们的报告首次直观呈现了使用左心室辅助装置患者的未闭卵圆孔解剖结构。