Danforth R A, Clark D E
Department of Oral and Maxillofacial Imaging, University of Southern California, Los Angeles, CA 90089-0641, USA.
Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2000 Feb;89(2):236-43. doi: 10.1067/moe.2000.103526.
The purpose of this investigation was to measure the tissue-absorbed dose and to calculate the whole-body effective dose (E) for a new generation panoramic machine (Planmeca PM 2002 CC Proline), operating in the panoramic examination mode. Comparisons could then be made with historical panoramic and intraoral radiographic dose measurement values and effective dose estimates.
Thermoluminescent dosimeters were embedded in a humanoid, tissue-equivalent phantom at anatomically significant sites, representing key tissues. Absorbed dose measurements were obtained after every 5 panoramic exposures of a 25-exposure total. The measured average tissue-absorbed doses from a single panoramic exposure were used in the calculation of the whole-body E.
The whole-body E for the PM 2002 CC Proline panoramic examination is 3.85 microSv. This is below the panoramic average of 6.7 microSv. The PM 2002 CC Proline panoramic examination delivers approximately 5% to 12% of the E of a complete mouth intraoral radiographic examination.
The effective dose from the PM 2002 CC Proline examination is at the low end of the range for other panoramic machines and is far below either a D-speed or E-speed film intraoral radiographic examination.