Angulation errors in parallel-hole and fanbeam collimators: computer controlled quality control and acceptance testing procedure.

作者信息

Eckholt M, Bergmann H

机构信息

Department of Biomedical Engineering and Physics, University Hospital, Vienna, Austria.

出版信息

J Nucl Med. 2000 Mar;41(3):548-55.

PMID:10716331
Abstract

UNLABELLED

Angulation errors in collimators of 1 degrees or even less can seriously diminish the resolution of SPECT images. We have developed a computer-controlled quality control procedure that can be used for acceptance testing and regular routine checks.

METHODS

Using a marker point source and a computer-controlled x-y positioning table, we investigated 7 parallel-hole and 3 fanbeam collimators. The results are presented as collimator surface maps, which are easy to interpret visually.

RESULTS

The measurement accuracy for absolute angulation errors was better than 0.32 degrees. Regional variations in channel tilt could be detected with an accuracy better than 0.16 degrees. Six parallel-hole collimators were found acceptable for high-resolution SPECT imaging. For a parallel-hole collimator that had to be replaced because of nonoptimal image quality, our measurements clearly identified regions of directionally uniform angulation errors. Two fanbeam collimators showed slight concavities.

CONCLUSION

Automation of the measurement and evaluation process make this procedure suitable for both acceptance tests and routine quality control checks. It can be applied to parallel-hole, fanbeam, converging, and diverging collimators, regardless of their individual geometry. No technical collimator specifications are needed. Our results reveal subtle mechanical deformations of collimators. They also show that for a detailed investigation, angulation error surface maps should be used to discover regional preferences in channel orientation.

摘要

相似文献

1
Angulation errors in parallel-hole and fanbeam collimators: computer controlled quality control and acceptance testing procedure.
J Nucl Med. 2000 Mar;41(3):548-55.
2
Measurement of collimator hole angulation and camera head tilt for slant and parallel hole collimators used in SPECT.
J Nucl Med. 1987 Oct;28(10):1592-8.
3
Half-fanbeam collimators combined with scanning point sources for simultaneous emission-transmission imaging.半扇形准直器与扫描点源相结合用于同时发射-透射成像。
J Nucl Med. 1998 Nov;39(11):1996-2003.
4
Combination of converging collimators for high-sensitivity brain SPECT.用于高灵敏度脑单光子发射计算机断层扫描的汇聚准直器组合。
J Nucl Med. 2009 Sep;50(9):1548-56. doi: 10.2967/jnumed.109.062653. Epub 2009 Aug 18.
5
Slit slat collimator optimization with respect to MTF.狭缝平板准直器关于调制传递函数的优化。
Appl Radiat Isot. 2005 Mar;62(3):461-8. doi: 10.1016/j.apradiso.2004.07.009.
6
Tomographic image quality of rotating slat versus parallel hole-collimated SPECT.旋转叶片与平行孔准直 SPECT 的断层图像质量。
Phys Med Biol. 2011 Nov 21;56(22):7205-22. doi: 10.1088/0031-9155/56/22/013. Epub 2011 Oct 25.
7
System characteristics of SPECT with a slat collimated strip detector.具有板条准直条形探测器的单光子发射计算机断层扫描(SPECT)系统特性
Phys Med Biol. 2006 Jan 21;51(2):391-405. doi: 10.1088/0031-9155/51/2/014. Epub 2006 Jan 4.
8
Collimator angulation error and its effect on SPECT.
J Nucl Med. 1990 May;31(5):655-9.
9
Collimators.准直器。
Q J Nucl Med. 2002 Mar;46(1):8-15.
10
The influence of collimators on SPECT center of rotation measurements: artifact generation and acceptance testing.准直器对单光子发射计算机断层扫描旋转中心测量的影响:伪影产生与验收测试
J Nucl Med. 1988 Aug;29(8):1393-7.