Microscopic analysis of three different spinal needle tips after experimental subarachnoid puncture.

作者信息

Puolakka R, Andersson L C, Rosenberg P H

机构信息

Department of Anaesthesiology, Helsinki University Central Hospital, Töölö Hospital, Finland.

出版信息

Reg Anesth Pain Med. 2000 Mar-Apr;25(2):163-9. doi: 10.1053/rapm.2000.0250163.

Abstract

BACKGROUND AND OBJECTIVES

Previous studies have shown the vulnerability of the tips of cutting type thin spinal needles and the possibility of foreign material passing into the subarachnoid space during the lumbar puncture. We made a microscopic analysis to compare two commonly used noncutting pencil-point spinal needles with different tip designs. Needles with a cutting tip design were included as reference.

METHODS

Four fresh cadavers were placed in the lateral position and their backs were scrubbed with disinfectant solution containing 0.1% fluorescein. Thirty-two spinal needles (27 gauge) of each type (modified Quincke, modified Sprotte, and modified Whitacre) were inserted through an introducer at interspaces L2-5 into the subarachnoid space. Under visual control (spinal canal opened ventrally) all the needle tips were cut after successful subarachnoid puncture; 16 needles of each tip design were investigated under a fluorescence microscope, and another 16 needle tips were collected into test tubes and cytocentrifuged smears were prepared. The tips and smears with the most obvious findings were photographed under a microscope.

RESULTS

On microscopy, only 2 needle tips were damaged (1 modified Quincke and 1 modified Whitacre). Visible fluorescent tissue particles were more frequently seen on modified Quincke needles (56%) compared with modified Sprotte (37%) and Whitacre (37%) needles (NS). In the cytocentrifugation smears, the largest clusters of epithelioid cells and muscle fibers were observed in the Quincke group. In the Whitacre group many fewer and smaller cell clusters including small muscle particles were seen, and only minor epithelioid cells were found in the Sprotte group.

CONCLUSIONS

Tissue coring seems to be a common phenomenon during lumbar puncture. The most prominent attachments appeared with a cutting Quincke-type spinal needle.