Krieglstein G K, Andrae K
Albrecht Von Graefes Arch Klin Exp Ophthalmol. 1975;193(2):145-52. doi: 10.1007/BF00419358.
The Armaly technique, a method of screening for glaucomatous field defects was evaluated in 100 normal eyes and 38 selected central field defects using the Goldmann perimeter and the Auto-Plot tangent screen. The screening results obtained from the Goldmann perimeter and the tangent screen using the selective perimetry of Armaly were compared with the results obtained from the Friedmann central field analyser. Testing normal eyes we found the highest initial false positive rate with the Friedmann analyser and the lowest initial false positive rate with the Auto-Plot tangent screen. The specificity of the Friedmann analyser for detecting early central field changes is less than the Goldmann perimeter or the Auto-Plot tangent screen when using Armaly's technique. On the other hand, the Friedmann analyser turned out to be the easiest instrument to handle and requires least experiences for the observer. Some problems of mass screening perimetry are discussed.
采用戈德曼视野计和自动绘图切线屏,对100只正常眼睛和38例选定的中心视野缺损进行了评估,以检测青光眼视野缺损的阿马利技术。将使用阿马利选择性视野检查法从戈德曼视野计和切线屏获得的筛查结果与从弗里德曼中心视野分析仪获得的结果进行比较。在测试正常眼睛时,我们发现弗里德曼分析仪的初始假阳性率最高,而自动绘图切线屏的初始假阳性率最低。当使用阿马利技术时,弗里德曼分析仪检测早期中心视野变化的特异性低于戈德曼视野计或自动绘图切线屏。另一方面,弗里德曼分析仪被证明是最易于操作的仪器,对观察者的经验要求最少。文中还讨论了群体筛查视野检查的一些问题。