Stelzer R
Albrecht Von Graefes Arch Klin Exp Ophthalmol. 1975;194(3):217-20. doi: 10.1007/BF00496878.
The applanation tonometry is influenced by several factors not considered by the law of Imbert-Fick. For instance, a change in the conditions of moistening the cornea may falsify the result of measuring. This has been proved clinically and experimentally with Polyvinylpyrrolidon. The fault in measuring is larger with applanation tonometers producing variable applanation areas along with constant pressure than with those flattening small invariable corneal area. Aside of the difficulty of exact measurement of the applanation areas the latter tonometers are superior to the former ones because of this reason, too.
压平眼压测量法受多种伊姆伯特-菲克定律未考虑的因素影响。例如,角膜湿润条件的变化可能会使测量结果失真。这已通过聚乙烯吡咯烷酮在临床和实验中得到证实。与那些使小的不变角膜区域变平的眼压计相比,产生可变压平面积且压力恒定的压平眼压计测量误差更大。除了精确测量压平面积存在困难外,后一种眼压计也因此优于前一种眼压计。