Quantification of D-dimer using a new fully automated assay: its application for the diagnosis of deep vein thrombosis.

作者信息

Villa P, Ferrando F, Serra J, Faus H, Mira Y, Vayá A, Aznar J

机构信息

Department of Clinical Pathology, La Fe University Hospital, Valencia Spain.

出版信息

Haematologica. 2000 May;85(5):520-4.

Abstract

BACKGROUND AND OBJECTIVE

A D-dimer assay can be helpful to rule out thromboembolism provided it is sensitive, reliable, fast and easy to perform. Tests based on the ELISA methodology have a high diagnostic sensitivity, and are therefore adequate for excluding deep venous thrombosis (DVT). The drawbacks are their long assay times, unsuitability to be run on single samples and cost. New methods have been developed, based either on the same principle, by immunofiltration or by microlatex immunoturbidimetric assays which seem to reach the high sensitivity and negative predictive value (NPV) required, but allowing fast and quantitative single sample analysis. The aim of this work was to evaluate one rapid test, a fully automated quantitative assay (IL Test D-dimer, run on an ACL 7000 coagulation analyzer, Instrumentation Laboratory).

DESIGN AND METHODS

We compared the diagnostic value of IL Test for DVT with that of an ELISA (Dimertest Gold EIA Agen Biomedical Limited, Acacia Ridge, Australia). Eighty-six patients (43 men, 43 women, mean age: 61 years) showing DVT symptoms formed the population for this non-randomized controlled trial in a referral center. The diagnosis of DVT based on the clinical history, was confirmed by serial compression ultrasonography (CUS) with Doppler flow in 62 patients.

RESULTS

The IL Test D-dimer proved to be rapid, automated and well suited for individual tests with a good reproducibility in three control plasmas with different concentrations of D-dimer (coefficient of variation range 0.54-3. 87%). Its performance was comparable to that of the Dimertest Gold EIA, as indicated by the areas under the receiver operating characteristic curves (Dimertest Gold EIA 0.748; IL Test D-dimer 0.70). On the basis of kappa coefficients, there was a good concordance between the Dimertest Gold EIA and IL Test D-dimer when the receiver operating characteristic (ROC) curves suggested cut-offs were used. The sensitivity (98.3%) and NPV (88.9%) shown by IL Test D-dimer are comparable or even better than those obtained for EIA (95%, 80%, respectively).

INTERPRETATION AND CONCLUSIONS

This study shows that the new method can be included in prospective clinical trials to test the utility of D-dimer measurement in combination with other non-invasive diagnostic procedures in the management of DVT diagnosis.

摘要

背景与目的

D - 二聚体检测若具备敏感性、可靠性、快速性及操作简便性,将有助于排除血栓栓塞。基于酶联免疫吸附测定(ELISA)方法的检测具有较高的诊断敏感性,因此适用于排除深静脉血栓形成(DVT)。其缺点在于检测时间长、不适用于单个样本检测以及成本高。基于相同原理,已开发出免疫过滤或微乳胶免疫比浊法等新方法,这些方法似乎能达到所需的高敏感性和阴性预测值(NPV),且能实现快速定量的单个样本分析。本研究的目的是评估一种快速检测方法,即一种全自动定量检测(IL Test D - 二聚体,在ACL 7000凝血分析仪上运行,仪器实验室)。

设计与方法

我们将IL Test D - 二聚体检测对DVT的诊断价值与ELISA(Dimertest Gold EIA,Agen Biomedical Limited,澳大利亚相思岭)的诊断价值进行了比较。86例出现DVT症状的患者(43例男性,43例女性,平均年龄:61岁)构成了该转诊中心这项非随机对照试验的研究对象。根据临床病史诊断为DVT的患者中,62例经连续压迫超声检查(CUS)及多普勒血流检查得以确诊。

结果

IL Test D - 二聚体检测结果显示其快速、自动化,非常适合单个检测,在三种不同浓度D - 二聚体的对照血浆中具有良好的重复性(变异系数范围为0.54 - 3.87%)。其性能与Dimertest Gold EIA相当,如受试者操作特征曲线下面积所示(Dimertest Gold EIA为0.748;IL Test D - 二聚体为0.70)。基于kappa系数,当使用受试者操作特征(ROC)曲线建议的临界值时,Dimertest Gold EIA与IL Test D - 二聚体之间具有良好的一致性。IL Test D - 二聚体显示的敏感性(98.3%)和NPV(88.9%)与EIA所获结果(分别为95%和80%)相当甚至更好。

解读与结论

本研究表明,这种新方法可纳入前瞻性临床试验,以测试D - 二聚体检测与其他非侵入性诊断程序联合用于DVT诊断管理的效用。