[Audit on pelvimetry at a polyvalent radiology department].

作者信息

Davy-Miallou C, Fraboulet P, Valéro M, Mourey I, Grellet J, Grenier P

机构信息

Service de Radiologie Polyvalente Diagnostique et Interventionnelle, Paris, France.

出版信息

J Radiol. 2000 May;81(5):529-34.

Abstract

PURPOSE

Study of the dose absorbed by the fetus during CT pelvimetry.

MATERIAL AND METHODS

Twenty consecutive pelvimetries on pregnant women showed the mean fetal location in regard of the maternal pelvic skeleton (retrospective study). Then, thermoluminescent radiation detectors were put in the antropomorph phantom slices, in housing correspondent to the fetal's volume. The location of this volume has been determined by the first study. Two different CT units were tested.

RESULTS

In case of cephalic presentation, the maximum fetal dose was 3 mGy. In case of pelvic presentation, because of an additional frontal topogram, the dose reached 3,3 to 3,6 mGy.

CONCLUSION

We recommend: 1) reduction of parameters (120kV and 100mAs) for "normal" women; using other parameters for larger women; interval follow-up to evaluate results using these parameters; 2) tilting the bisciatic plane of section so to avoid direct exposure to the fetus; 3) limitation to 25 cm for the lateral scout; suppression of the routine frontal scout in case of pelvic presentation.