Factors influencing the output of an implantable force transducer.

作者信息

Fleming B C, Peura G D, Beynnon B D

机构信息

Department of Orthopaedics and Rehabilitation, McClure Musculoskeletal Research Center, University of Vermont, Burlington, VT 05406-0084, USA.

出版信息

J Biomech. 2000 Jul;33(7):889-93. doi: 10.1016/s0021-9290(00)00019-1.

Abstract

The objective of this study was to evaluate the performance of the Arthroscopically Implantable Force Probe (AIFP; MicroStrain, Burlington VT) for measuring force in a patellar tendon graft. Transducer drift, reproducibility of output due to the number of loading cycles and device location, and sensitivity to the tendon cross-sectional area were investigated. The AIFP was initialized, and then implanted into five human patellar tendon grafts three times; twice within the same location and once in a different location. The tendons were cyclically loaded in uniaxial tension for 500 cycles in each insertion site. The AIFP was then removed from the tendon and the baseline output was remeasured. It was determined that transducer drift was negligible. The relationship between the tensile load applied to the graft and AIFP output was quadratic and specimen dependent. The cyclic load response of the tendon-AIFP interface demonstrated a 24.9% decrease over the first 20 loading cycles, and subsequent cycling yielded relatively reproducible output. The output of the transducer varied when it was removed from the tendon and then reimplanted in the same location (range 3.7-109. 4% error), as well as in the second location (range 1.5-202.8% error). No correlation was observed between the cross-sectional area of the tendon and transducer output. This study concludes that implantable force probes should be used with caution and calibrated without removing the transducer from the graft.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验