Takayama T, Nakamura N, Motomura N, Mori I, Ozaki T, Ohno R
Medical Systems Research & Development Center, Toshiba Corporation Medical Systems Company.
Kaku Igaku. 2000 May;37(3):181-7.
To evaluate the performance of a semiconductor detector for use in a gammacamera system, we assembled a detector with a small field of view--1 inch x 1 inch and 1 inch x 2 inch--made from CdTe (Cadmium telluride). We then compared the planar images and energy resolution of the resulting detectors against those of a conventional gammacamera. Pixel pitch of the detector was 1.6 mm x 1.6 mm, and was manufactured by Acrorad Corporation. Average FWHM of the energy spectrum for the semiconductor detector was 5.11% (SD: 0.80%, Best: 3.26%, Worst: 6.68%). The planar images obtained were of a letter phantom made from pieces of lead and of an IMP brain phantom. Since the field of view of the semiconductor detector was small, the image of the IMP brain phantom was acquired by moving the semiconductor over the collimated detector module until the area of the entire phantom was covered. The images from the semiconductor assembly were compared with those from a conventional gammacamera using the same conditions, and it was found that visual image quality was superior to those of the conventional camera system.
为评估用于伽马相机系统的半导体探测器的性能,我们组装了一个小视野探测器——1英寸×1英寸和1英寸×2英寸——由碲化镉(CdTe)制成。然后我们将所得探测器的平面图像和能量分辨率与传统伽马相机的进行比较。探测器的像素间距为1.6毫米×1.6毫米,由Acrorad公司制造。半导体探测器能谱的平均半高宽为5.11%(标准差:0.80%,最佳:3.26%,最差:6.68%)。所获得的平面图像是由铅片制成的字母体模和一个IMP脑体模的图像。由于半导体探测器的视野较小,通过在准直探测器模块上移动半导体来获取IMP脑体模的图像,直到覆盖整个体模区域。在相同条件下,将半导体组件的图像与传统伽马相机的图像进行比较,发现视觉图像质量优于传统相机系统。