使用Acs:180血清皮质醇化学发光免疫分析法测定尿游离皮质醇。

Measurement of urinary free cortisol using the Acs:180 serum cortisol chemiluminescent immunoassay.

作者信息

Kennedy D M, Selby C, Lawson N

机构信息

Department of Clinical Chemistry, Nottingham City Hospital NHS Trust, UK.

出版信息

Ann Clin Biochem. 2000 Jul;37 ( Pt 4):520-8. doi: 10.1177/000456320003700414.

Abstract

Measurement of urinary free cortisol using the Bayer Automated Chemiluminescent System (ACS:180 PLUS) was evaluated and compared with an in-house extraction radioimmunoassay (RIA). Inter-assay coefficients of variation were acceptable, being respectively 5.3%, 4.8% and 3.8% at 141, 406 and 942 nmol/L (n=20) for the ACS direct assay and 19.4% and 12.1% at 27 and 116 nmol/L with dichloromethane extraction (n=10). Using tritiated cortisol, mean extraction efficiency for the ACS extraction protocol was 89% compared with 104% for the RIA method (n=6, P<0.001) and using urine spiked with ACS serum calibrator, extraction efficiency was only 76% (standard deviation 8.8%, n=49) on the ACS. Urine spiked with cortisol dissolved in ethanol also gave significantly lower recoveries on the ACS for both direct and extraction methods compared with the RIA. Regression analysis of results from a mixture of control serum samples and samples from patients (n=93) showed good correlations between the direct and extraction ACS methods and the RIA extraction assay. The median concentration in 23 normal subjects and 95% reference intervals (nmol/24 h) were: ACS direct, 237 (135-505); ACS extraction, 91 (33-239); and RIA extraction, 146 (80-334). 5Beta-dihydrocortisol was the only major interferent in all assays. The ACS extraction assay showed acceptable performance and correlated well with the extraction RIA, although there was evidence of matrix-dependent effects causing low recovery.

摘要

使用拜耳自动化学发光系统(ACS:180 PLUS)对尿游离皮质醇进行测量,并与内部萃取放射免疫分析法(RIA)进行比较。批间变异系数是可接受的,对于ACS直接测定法,在141、406和942 nmol/L(n = 20)时分别为5.3%、4.8%和3.8%,对于二氯甲烷萃取法,在27和116 nmol/L(n = 10)时分别为19.4%和12.1%。使用氚化皮质醇,ACS萃取方案的平均萃取效率为89%,而RIA法为104%(n = 6,P<0.001),并且使用添加了ACS血清校准品的尿液时,ACS上的萃取效率仅为76%(标准差8.8%,n = 49)。添加溶解在乙醇中的皮质醇的尿液,与RIA相比,ACS直接法和萃取法的回收率也显著降低。对对照血清样本和患者样本(n = 93)的混合物结果进行回归分析表明,ACS直接法和萃取法与RIA萃取法之间具有良好的相关性。23名正常受试者的中位数浓度和95%参考区间(nmol/24 h)为:ACS直接法,237(135 - 505);ACS萃取法,91(33 - 239);RIA萃取法,146(80 - 334)。5β - 双氢皮质醇是所有测定中唯一的主要干扰物。ACS萃取法表现出可接受的性能,并且与萃取RIA法相关性良好,尽管有证据表明基质依赖性效应导致回收率较低。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验