Jørgensen J T, Andersen P B, Rosholm A, Bjarnason N H
Pronosco A/S, Torsana Park, DK-2950 Vedbaek, Denmark.
Clin Physiol. 2000 Sep;20(5):330-5. doi: 10.1046/j.1365-2281.2000.00268.x.
The precision of any given method for measurement of bone mineral density (BMD) is important in relation to the interpretation of repeated measurements over time, e.g. to monitor the course of suspected osteoporosis or follow the effect of therapy. In the present study a new bone densitometer using the digital X-ray radiogrammetry (DXR) method (Pronosco X-posure Systemtrade mark) is investigated with respect to its short-term precision. The study was carried out on two groups of females, one consisting of 20 women between the ages of 30 and 40, and the other of 20 post-menopausal women above the age of 64. The mean age of the premenopausal women was 35.2 years and the mean DXR BMD was 0.578 g cm-2. The mean age of the post-menopausal women was 68.2 years and the mean DXR BMD was 0.489 g cm-2. The short-term precision of the two groups was evaluated using the coefficient of variation (CV%) and corresponding 90% confidence intervals. The coefficient of variation in the premenopausal group was 0.68% with a 90% confidence interval of 0. 57%-0.83%. The coefficient of variation in the postmenopausal group was 0.61% with a 90% confidence interval of 0.52-0.75%. It can be concluded from the present study that the short-term in vivo precision error of the DXR method is low in both pre- and post-menopausal women. When the results of the study are compared to data reported in the literature, the performance of the DXR method seems to be at least equivalent with peripheral DXA.
任何给定的骨密度(BMD)测量方法的精度对于解释随时间的重复测量结果都很重要,例如监测疑似骨质疏松症的病程或跟踪治疗效果。在本研究中,对一种采用数字X射线摄影测量法(DXR)的新型骨密度仪(Pronosco X-posure System商标)的短期精度进行了研究。该研究针对两组女性进行,一组由20名年龄在30至40岁之间的女性组成,另一组由20名64岁以上的绝经后女性组成。绝经前女性的平均年龄为35.2岁,平均DXR骨密度为0.578 g/cm²。绝经后女性的平均年龄为68.2岁,平均DXR骨密度为0.489 g/cm²。使用变异系数(CV%)和相应的90%置信区间评估两组的短期精度。绝经前组的变异系数为0.68%,90%置信区间为0.57%-0.83%。绝经后组的变异系数为0.61%,90%置信区间为0.52-0.75%。从本研究可以得出结论,DXR方法在绝经前和绝经后女性中的短期体内精度误差都很低。当将该研究结果与文献报道的数据进行比较时,DXR方法的性能似乎至少与外周双能X线吸收法相当。