Pfeil Alexander, Renz Diane M, Fröber Rosemarie, Hansch Andreas, Lehmann Gabriele, Sommerfeld Julia, Malich Ansgar, Wolf Gunter, Böttcher Joachim
Department of Internal Medicine III, Jena University Hospital - Friedrich Schiller University Jena, Erlanger Allee 101, 07747 , Jena, Germany,
Int J Comput Assist Radiol Surg. 2015 May;10(5):587-92. doi: 10.1007/s11548-014-1076-y. Epub 2014 Jul 3.
Digital X-ray radiogrammetry (DXR) is a computer-assisted technique used to quantify cortical bone density of the metacarpals. The influence of metacarpal bone rotation and type of cast material on bone mineral density (BMD) measurements using the DXR technique was tested.
The bone mineral density of the hand was measured by DXR, and rotation error (DXR-RE) as coefficients of variation were calculated, to verify reliability and reproducibility of this radiogeometric technique to assess in particular minor disease-related changes in the metacarpal bone mass. The reproducibility of the DXR measurements was also investigated using different cast materials (mull, elastic, and plastic).
There were no significant changes in absolute values of DXR-BMD observed between 0 to [Formula: see text] angulation. The relative DXR-RE ranged between 0 % (degree 1) and 0.70 % (degrees 15 and 19) for DXR-BMD. Regarding the different cast materials, DXR-BMD revealed a coefficient of variation with 0.41 % (mull cast) and 0.21 % (elastic cast). For the plastic cast, the DXR technique was not able to perform an analysis of DXR-BMD.
The study revealed no significant influence of metacarpal rotation on the measurements of metacarpal bone mineral density as estimated by DXR. DXR measurements are not optimal when cast material is used. DXR can accurately quantify periarticular cortical bone mass. This is significant especially for rheumatoid arthritis and related conditions where X-ray imaging of arthritic hands with varying degrees of deformity is performed.
数字X线摄影测量法(DXR)是一种用于量化掌骨皮质骨密度的计算机辅助技术。测试了掌骨旋转和石膏材料类型对使用DXR技术测量骨矿物质密度(BMD)的影响。
采用DXR测量手部骨矿物质密度,并计算作为变异系数的旋转误差(DXR-RE),以验证这种放射几何学技术评估掌骨骨量中特别是微小疾病相关变化的可靠性和可重复性。还使用不同的石膏材料(纱布、弹性材料和塑料)研究了DXR测量的可重复性。
在0至[公式:见正文]度成角之间,未观察到DXR-BMD绝对值有显著变化。DXR-BMD的相对DXR-RE在0%(1度)至0.70%(15度和19度)之间。对于不同的石膏材料,DXR-BMD的变异系数为0.41%(纱布石膏)和0.21%(弹性石膏)。对于塑料石膏,DXR技术无法对DXR-BMD进行分析。
该研究表明,掌骨旋转对DXR估计的掌骨骨矿物质密度测量无显著影响。使用石膏材料时,DXR测量并非最佳。DXR可准确量化关节周围皮质骨量。这对于类风湿关节炎及相关疾病尤其重要,因为在这些疾病中需要对不同程度畸形的手部进行X线成像。