Hill D M, Barnes R K, Wong H K, Zawadzki A W
Australian Nuclear Science and Technology Organisation Menai, NSW.
Appl Radiat Isot. 2000 Sep;53(3):415-9. doi: 10.1016/s0969-8043(99)00280-8.
Previous analytical methods employed to measure 99Tc:99mTc ratios in chromatographic 99mTc generators, derived from fission-based 99Mo, were limited in monitoring the ratios in low activity generators, over the extent of their elution regimes. For regulatory purposes, ICP-MS was employed to measure total technetium in generator eluates down to a level of 200 pg ml(-1) (200 ppt). Measured 99Tc:99mTc ratios compared very well with the theoretically computed values, ranging from 47:1 for a first elution of a generator with a long ingrowth time to 1.4:1 for the 17th elution of a generator with a short ingrowth time.
以往用于测量基于裂变产生的99Mo的色谱法99mTc发生器中99Tc:99mTc比率的分析方法,在监测低活度发生器洗脱范围内的比率时存在局限性。出于监管目的,采用电感耦合等离子体质谱法(ICP-MS)测量发生器洗脱液中的总锝,检测下限可达200 pg ml(-1)(200 ppt)。测得的99Tc:99mTc比率与理论计算值非常吻合,范围从生长时间长的发生器首次洗脱时的47:1到生长时间短的发生器第17次洗脱时的1.4:1。