Dammann F, Bode A, Heuschmid M, Kopp A, Georg C, Pereira P L, Claussen C D
Abt. f. Radiologische Diagnostik, Radiologische Universitätsklinik Tübingen.
Rofo. 2000 Aug;172(8):701-6. doi: 10.1055/s-2000-7175.
To evaluate the potential of multislice CT in the diagnosis of the paranasal sinuses.
Dose measurements were performed exposing an Alderson Rando phantom in a four-slice spiral CT (MS-CT) while applying a variety of scan parameters. Additionally, 30 consecutive patients underwent a transversal examination by the MS-CT using 1/1/3.5/0.5 mm spiral parameters and an additional transversal or coronal scan on a conventional single slice spiral CT (SS-CT) with 2/3/1 mm. Coronal reformations of the MS-CT were compared with the primary coronal SS-CT, or coronal reformations of the transversal SS-CT, respectively, with regard to image quality and depiction of relevant anatomical details of the region.
Superficial exposure values at the level of the eye lenses as well as for the thyroid gland were superior for MS-CT (3.62 mGy, and 0.12 mGy, resp.) as compared to SS-CT (2.96 mGy, and 0.07 mGy). Image quality was equal or superior for MS-CT as compared to SS-CT in all but one case. Drawbacks of SS-CT, such as dental amalgam artifacts, stair step artifacts or partial volume artifacts did not notably affect the coronal reformations of MS-CT.
MS-CT seems to have the potential to replace primary coronal CT of the paranasal sinuses without any loss of image quality, but may even improve the overall diagnostic value. Radiation doses may still have to be reduced.
评估多层螺旋CT在诊断鼻窦方面的潜力。
在四排螺旋CT(MS-CT)中对Alderson Rando体模进行剂量测量,同时应用各种扫描参数。此外,30例连续患者接受了MS-CT的横断面检查,使用1/1/3.5/0.5 mm螺旋参数,并在传统单层螺旋CT(SS-CT)上进行了额外的横断面或冠状面扫描,参数为2/3/1 mm。将MS-CT的冠状面重建分别与原发性冠状面SS-CT或横断面SS-CT的冠状面重建在图像质量和该区域相关解剖细节的显示方面进行比较。
与SS-CT(分别为2.96 mGy和0.07 mGy)相比,MS-CT在晶状体水平以及甲状腺的表面照射值更高(分别为3.62 mGy和0.12 mGy)。除1例情况外,MS-CT的图像质量与SS-CT相当或更优。SS-CT的缺点,如牙科汞合金伪影、阶梯状伪影或部分容积伪影,对MS-CT的冠状面重建没有明显影响。
MS-CT似乎有潜力在不损失任何图像质量的情况下取代鼻窦原发性冠状面CT,甚至可能提高整体诊断价值。辐射剂量可能仍需降低。