Suppr超能文献

基于体模研究的磁共振成像中几何畸变的可重复性

Reproducibility of geometric distortion in magnetic resonance imaging based on phantom studies.

作者信息

Mizowaki T, Nagata Y, Okajima K, Kokubo M, Negoro Y, Araki N, Hiraoka M

机构信息

Department of Radiology, Faculty of Medicine, Kyoto University, 54 Shogoin Kawahara-cho, Sakyo-ku, Kyoto 606-8507, Japan.

出版信息

Radiother Oncol. 2000 Nov;57(2):237-42. doi: 10.1016/s0167-8140(00)00234-6.

Abstract

BACKGROUND AND PURPOSE

Image distortion is one of the major drawbacks of magnetic resonance (MR) imaging for use in radiotherapy treatment planning (RTTP). In this study, the reproducibility of MR imaging distortion was evaluated by repeated phantom measurements.

MATERIALS AND METHODS

A grid-pattern acrylic phantom was scanned with a 0.2-Tesla permanent magnetic unit. We repeated a series of scans three times to evaluate the reproducibility of the distortion. In each series, co-ordinates at 432 intersections of the grid were measured for both T1- and T2-weighted spin-echo (SE) pulse sequences. Positional displacements and their variations at the intersections were calculated.

RESULTS

Averages of the displacements were distributed between 1.58 and 1.74 mm, and maximum values (MAX) between 12.6 and 15.0 mm. Within 120 mm of the image center, the average values ranged from 0.73 to 0.80 mm, and from 3.4 to 5.0 mm for MAX. The absolute values of the positional variations among three series were distributed between 0.41 and 0.88 mm for average values, and between 1.4 and 4.5 mm for MAX.

CONCLUSIONS

The positional variations were mostly within 3 pixels, and most of the positional displacements within the radius of 120 mm of the image center were 2 mm or less. Therefore, it will be possible to use this MR system in RTTP under limited situations, although careful applications are required for RTTP of the body. The development of a computer program to correct image distortion is expected.

摘要

背景与目的

图像畸变是磁共振(MR)成像应用于放射治疗计划(RTTP)的主要缺点之一。在本研究中,通过重复体模测量评估了MR成像畸变的可重复性。

材料与方法

使用0.2特斯拉永磁单元对网格图案的丙烯酸体模进行扫描。我们将一系列扫描重复三次以评估畸变的可重复性。在每个系列中,针对T1加权和T2加权自旋回波(SE)脉冲序列测量网格432个交点处的坐标。计算交点处的位置位移及其变化。

结果

位移平均值分布在1.58至1.74毫米之间,最大值(MAX)在12.6至15.0毫米之间。在图像中心120毫米范围内,平均值范围为0.73至0.80毫米,MAX范围为3.4至5.0毫米。三个系列之间位置变化的绝对值,平均值分布在0.41至0.88毫米之间,MAX分布在1.4至4.5毫米之间。

结论

位置变化大多在3个像素以内,图像中心半径120毫米范围内的大多数位置位移为2毫米或更小。因此,尽管在身体的RTTP中需要谨慎应用,但在有限情况下可以使用该MR系统进行RTTP。预计将开发一种用于校正图像畸变的计算机程序。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验