Mizowaki T, Nagata Y, Okajima K, Murata R, Yamamoto M, Kokubo M, Hiraoka M, Abe M
Department of Radiology, Faculty of Medicine, Kyoto University, Japan.
Radiology. 1996 Jun;199(3):855-60. doi: 10.1148/radiology.199.3.8638017.
To evaluate the geometric distortion on magnetic resonance (MR) images obtained with a permanent magnet system and determine the usefulness of MR imaging-assisted x-ray simulation in radiation therapy treatment planning (RTTP).
The authors measured the distortion on MR images of grid-pattern phantoms. MR imaging-assisted x-ray simulation was performed with skin markers in 14 patients with bone tumors. Treatment planning had already been performed with a conventional system.
On phantom images, most of the positional displacements within a 120-mm radius from the center of the static magnetic field were less than 2 mm; larger displacements were observed in the peripheral region of the images. MR imaging was useful in the RTTP of all patients. The original radiation field was modified after MR examination in six patients.
The amount of image distortion within the practical area is acceptable for RTTP. MR imaging-assisted x-ray simulation is useful for patients with bone tumors and warrants further investigation.
评估使用永磁体系统获得的磁共振(MR)图像上的几何畸变,并确定MR成像辅助X射线模拟在放射治疗治疗计划(RTTP)中的实用性。
作者测量了网格图案体模的MR图像上的畸变。对14例骨肿瘤患者使用皮肤标记进行了MR成像辅助X射线模拟。治疗计划已经使用传统系统完成。
在体模图像上,距静磁场中心半径120毫米范围内的大多数位置位移小于2毫米;在图像的周边区域观察到较大的位移。MR成像对所有患者的RTTP均有用。6例患者在MR检查后修改了原始辐射野。
实际区域内的图像畸变程度对于RTTP是可接受的。MR成像辅助X射线模拟对骨肿瘤患者有用,值得进一步研究。