In 23 patients, percutaneous atherectomy was performed with an endomyocardial biopsy device in 27 eccentric stenoses of lower limb arteries. The mean percentage stenosis was reduced from 82% to 14% (P: <.001). Histopathologic examination revealed no media or adventitial tissue in the extracted particles. The technical success and complication rates were similar to those reported for conventional percutaneous transluminal angioplasty of short symmetric lesions.
在23例患者中,使用心内膜心肌活检装置对27处下肢动脉偏心性狭窄进行了经皮斑块旋切术。平均狭窄百分比从82%降至14%(P<0.001)。组织病理学检查显示,提取的颗粒中无中膜或外膜组织。技术成功率和并发症发生率与报道的短段对称性病变传统经皮腔内血管成形术相似。