Ronco C, Brendolan A, Crepaldi C, Rodighiero M, Everard P, Ballestri M, Cappelli G, Spittle M, La Greca G
Department of Nephrology, St. Bortolo Hospital, Vicenza, Italy.
Int J Artif Organs. 2000 Sep;23(9):601-9.
The efficiency of a hemodialyzer is largely dependent on its ability to facilitate diffusion, since this is the main mechanism by which small solutes are removed. The diffusion process can be impaired if there is a mismatch between blood and dialysate flow distribution in the dialyzer. The objective of the paper was to study the impact of different dialysate compartment designs on dialysate flow distribution and urea clearances. Eighteen hollow fiber 1.3 m2 hemodialyzers were studied, 6 each of 3 designs: Type A--standard fiber bundle (PAN 65DX Asahi Medical, Tokyo, Japan); Type B--spacing filaments external to the fibers (PAN 65SF Asahi Medical, Tokyo, Japan); Type C--fibers waved to give Moiré structure (FB130 Nissho-Nipro, Osaka, Japan).
3 dialyzers of each type were studied following dye injection into the dialysate compartment. Dynamic sequential imaging of longitudinal sections of the dialyzer were undertaken, using a new generation helical CT scanner (X-Press/HS1 Toshiba Corporation, Tokyo, Japan). In vivo studies: 3 dialyzers of each type were studied, in randomized sequence, in 3 different patients under standardized dialysis conditions. Blood- and dialysate-side urea clearances were measured at 30 and 150 minutes of treatment. Macroscopic and densitometrical analysis revealed that flow distribution was most homogeneous in the dialyzer with Moiré structure (Type C) and least homogeneous in the standard dialyzer (Type A). Space yarns (Type B) gave an intermediate dialysate flow distribution. Significantly increased urea clearances (p<0.001) were seen with Types B and C, compared to the standard dialyzer. Type C (Moiré) had the highest clearances although these were not significantly greater than Type B (space yarns). In conclusion, more homogeneous dialysate flow distribution and improved small solute clearances can be achieved by use of spacing yarns or waved (Moiré structure) patterns of fiber packing in the dialyzer. These effects are achieved probably as a result of reduced dialysate channeling resulting in a lower degree of mismatch between blood and dialysate flows. The new radiological technique using the helical CT scanner allows detailed flow distribution analysis and has the potential for testing future modifications to dialyzer design.
血液透析器的效率很大程度上取决于其促进扩散的能力,因为这是清除小溶质的主要机制。如果透析器中血液和透析液的流量分布不匹配,扩散过程可能会受到损害。本文的目的是研究不同透析液腔室设计对透析液流量分布和尿素清除率的影响。研究了18个中空纤维1.3平方米的血液透析器,3种设计各6个:A型——标准纤维束(日本东京旭化成医疗株式会社的PAN 65DX);B型——纤维外部有间隔丝(日本东京旭化成医疗株式会社的PAN 65SF);C型——纤维呈波浪状形成莫尔条纹结构(日本大阪日商日东纺绩株式会社的FB130)。
向每种类型的3个透析器的透析液腔室注入染料后进行研究。使用新一代螺旋CT扫描仪(日本东京东芝公司的X-Press/HS1)对透析器的纵向切片进行动态序列成像。体内研究:每种类型的3个透析器在标准化透析条件下,以随机顺序在3名不同患者中进行研究。在治疗30分钟和150分钟时测量血液和透析液侧的尿素清除率。宏观和密度分析显示,具有莫尔条纹结构的透析器(C型)中流量分布最均匀,标准透析器(A型)中最不均匀。间隔丝(B型)的透析液流量分布处于中间水平。与标准透析器相比,B型和C型的尿素清除率显著提高(p<0.001)。C型(莫尔条纹)的清除率最高,尽管与B型(间隔丝)相比没有显著更高。总之,通过在透析器中使用间隔丝或纤维堆积的波浪状(莫尔条纹结构)模式,可以实现更均匀的透析液流量分布和改善小溶质清除率。这些效果可能是由于透析液通道减少,导致血液和透析液流量之间的不匹配程度降低。使用螺旋CT扫描仪的新放射技术允许进行详细的流量分布分析,并有可能测试未来透析器设计的改进。