Jacobson K, Thompson A, Browne G, Shasserre C, Seelig S A, King W
Assay Development, Vysis, Inc, Downers Grove, IL 60515, USA.
Mol Diagn. 2000 Sep;5(3):209-20. doi: 10.1054/modi.2000.9731.
In fluorescence in situ hybridization (FISH) applications, the efficiency of probe hybridization is greatly enhanced by treating the cell or tissue preparation with a variety of reagents that make the target permeable while preserving important morphological features. Pretreatment protocols can be very labor intensive, adding cost to the test. The automation of specimen pretreatment eliminates human variation in the procedure through standardization of such variables as treatment times and temperatures.
We developed an instrument, the VP 2000 Processor, that can process up to 50 specimens simultaneously under computer control. In this comparative FISH study of matched specimens, one set was processed according to the manual pretreatment protocol and compared with specimens processed by the VP 2000. Processed specimen types included paraffin-embedded breast tissue, uncultured amniocytes, bone marrow, peripheral-blood lymphocytes, and uroepithelial cells recovered from urine. Data show equivalent or brighter and more specific overall signal quality compared with matched manually executed controls.
The automation of sample pretreatment for FISH provides a superior, more consistent level of performance than the manual format.
在荧光原位杂交(FISH)应用中,通过用多种试剂处理细胞或组织标本,可在保留重要形态特征的同时使靶标具有通透性,从而大大提高探针杂交效率。预处理方案可能非常耗费人力,增加检测成本。标本预处理的自动化通过对处理时间和温度等变量进行标准化,消除了该过程中的人为差异。
我们开发了一种仪器,即VP 2000处理器,它可以在计算机控制下同时处理多达50个标本。在这项对匹配标本的比较FISH研究中,一组标本按照手动预处理方案进行处理,并与由VP 2000处理的标本进行比较。处理的标本类型包括石蜡包埋的乳腺组织、未培养的羊膜细胞、骨髓、外周血淋巴细胞以及从尿液中回收的尿上皮细胞。数据显示,与匹配的手动操作对照相比,整体信号质量相当或更亮且更特异。
FISH样本预处理的自动化提供了比手动方式更优越、更一致的性能水平。