Dose-volume histograms computation comparisons using conventional methods and optimized fast Fourier transforms algorithms for brachytherapy.

作者信息

Kemmerer T, Lahanas M, Baltas D, Zamboglou N

机构信息

Department of Medical Physics & Engineering, Strahlenklinik, Klinikum Offenbach, Germany.

出版信息

Med Phys. 2000 Oct;27(10):2343-56. doi: 10.1118/1.1312810.

Abstract

In anatomy based optimization procedures for large volume implants the calculation of dose-volume histograms (DVH) accounts for the major part of the time involved and can be as long as a few hours. This time is proportional to the number of seeds or source dwell positions required for the implant. A procedure for the calculation of brachytherapy seed dose distribution calculation employing fast Fourier transforms (FFT) and the convolution theorem has been described by others and was supposed to significantly improve the speed of the dose distribution computation. Using new significantly improved FFT algorithms and various other optimization techniques we have compared the calculated differential and integral DVHs in high dose rate (HDR) brachytherapy with a single stepping source using actual clinical implants. This is so that we could assess the efficiency and accuracy of the FFT method with that of conventional methods. Our results showed that the FFT based method of calculating DVHs in brachytherapy is comparable in speed with conventional dose calculation methods, but only for implants with more than 287 sources. It is therefore of limited practical use even for large implants. This result is in direct opposition to the claim by other authors.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验