Honda K, Larheim T A, Johannessen S, Arai Y, Shinoda K, Westesson P L
Department of Radiology, Nihon University School of Dentistry, Tokyo, Japan.
Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2001 Feb;91(2):239-43. doi: 10.1067/moe.2001.111942.
The purpose of this study was to present a newly developed ortho cubic super-high resolution computerized radiographic technique (Ortho-CT) and its application to the temporomandibular joint.
A prototype was assembled on a Scanora (Soredex Findent Co) with the use of a 4-inch image intensifier instead of film. Data were collected from a single 360 degrees scan and a cylinder 32 mm in height and 38 mm in diameter. Images were reconstructed with a software program on a personal computer. Imaging data consisted of 240 (height) x 280 (diameter) cubic voxels, each with a dimension of 0.136 mm. With this small voxel size, the image resolution was high and was the same in any direction. Three patients with temporomandibular joint disease (trauma, pain and dysfunction, fibroosseous ankylosis) were evaluated with Ortho-CT, and the images were compared to routine radiographic films.
Preliminary clinical experience with Ortho-CT demonstrates subjectively superior image quality compared to panoramic and conventional tomographic images. The images were similar to conventional computed tomography, but this new radiographic technique is less expensive, takes less space, and gives a lesser patient dose.
本研究旨在介绍一种新开发的正立方超高分辨率计算机放射成像技术(Ortho-CT)及其在颞下颌关节中的应用。
在Scanora(Soredex Findent公司)上组装了一个原型,使用4英寸影像增强器而非胶片。数据来自单次360度扫描以及一个高32毫米、直径38毫米的圆柱体。图像通过个人计算机上的软件程序重建。成像数据由240(高)×280(直径)个立方体素组成,每个体素尺寸为0.136毫米。由于体素尺寸小,图像分辨率高且在任何方向上相同。对3例颞下颌关节疾病(创伤、疼痛和功能障碍、纤维性骨强直)患者进行了Ortho-CT评估,并将图像与常规放射照片进行比较。
Ortho-CT的初步临床经验表明,与全景和传统断层图像相比,其主观图像质量更优。图像与传统计算机断层扫描相似,但这种新的放射成像技术成本更低、占用空间更小且患者剂量更低。