A comparative study on the behavior of three different automatic mode switching dual chamber pacemakers to intracardiac recordings of clinical atrial fibrillation.

作者信息

Leung S K, Lau C P, Lam C T, Tse H F, Tang M O, Chung F, Ayers G

机构信息

Department of Medicine, Kwong Wah Hospital, Hong Kong.

出版信息

Pacing Clin Electrophysiol. 2000 Dec;23(12):2086-96. doi: 10.1111/j.1540-8159.2000.tb00781.x.

DOI:10.1111/j.1540-8159.2000.tb00781.x
PMID:11202252
Abstract

Automatic mode switching (AMS) allows patients with dual chamber pacemakers who develop paroxysmal AF to have a controlled ventricular rate. The aim of this study was to (1) compare the rate-controlled behavior of three AMS algorithms in response to AF, in terms of speed and stability of response and resynchronization to sinus rhythm, and (2) compare the influence of pacemaker programming on optimal mode switching. We studied 17 patients (12 men, 5 women; mean age 59 +/- 15 years) who developed AF during electrophysiological study. Unfiltered bipolar atrial electrograms during sinus rhythm and AF were recorded onto high fidelity tapes and replayed into the atrial port of three dual chamber pacemakers with different mode switching algorithms (Thera, Marathon, Meta). The Thera pacemaker uses rate smoothing, and mode switches occur when mean sensed atrial rate exceeds the predefined AMS rate (MR). Marathon mode switches after a programmable number of consecutive rapid atrial events (NR). Meta DDDR monitors the atrial rate by a counter for atrial cycles faster than the programmed AMS rate. It increases or decreases the counter if the atrial cycle length is shorter or longer than the programmed AMS interval, respectively. Mode switch occurs when the AF detection criteria are met (CR). A total of 260 rhythms were studied. NR was significantly faster than MR and CR (latency 2.5 +/- 3 s vs 26 +/- 7 s vs 15 +/- 22 s, respectively, P < 0.0001). During sustained AF, MR resulted in the most stable and regular ventricular rhythm compared to NR or CR. In CR, ventricular rate oscillated between AMS and atrial tracking (cycle length variations: 44 +/- 2 s vs 346 +/- 109 s vs 672 +/- 84 s, P < 0.05). At resumption of sinus rhythm, MR resynchronized after 143 +/- 22 s versus 3.4 +/- 0.7 s for NR and 5.9 +/- 1.1 s for CR, resulting in long periods of AV dissociation when a VVI/VVIR mode is used after AMS. Programming of atrial refractory periods did not affect AMS response, although the speed of AMS onset can be adjusted by programming of onset criteria in the Meta DDDR. AMS algorithms differ in their ability to handle recorded clinical atrial arrhythmias. The rapid-responding algorithm exhibits rate instability, whereas slow responding algorithm shows a long delay in response and risk of AV dissociation. Thus different instrumentation of AMS may have clinical implications in patients with dual chamber pacemakers who develop AF.

摘要

相似文献

1
A comparative study on the behavior of three different automatic mode switching dual chamber pacemakers to intracardiac recordings of clinical atrial fibrillation.
Pacing Clin Electrophysiol. 2000 Dec;23(12):2086-96. doi: 10.1111/j.1540-8159.2000.tb00781.x.
2
Programmed atrial sensitivity: a critical determinant in atrial fibrillation detection and optimal automatic mode switching.程控心房感知:心房颤动检测及最佳自动模式转换的关键决定因素。
Pacing Clin Electrophysiol. 1998 Nov;21(11 Pt 2):2214-9. doi: 10.1111/j.1540-8159.1998.tb01155.x.
3
Is automatic mode switching effective for atrial arrhythmias occurring at different rates? A study of the efficacy of automatic mode and rate switching to simulated atrial arrhythmias by chest wall stimulation.
Pacing Clin Electrophysiol. 2000 May;23(5):824-31. doi: 10.1111/j.1540-8159.2000.tb00850.x.
4
Automatic mode switching of implantable pacemakers: I. Principles of instrumentation, clinical, and hemodynamic considerations.植入式起搏器的自动模式切换:I. 仪器原理、临床及血流动力学考量
Pacing Clin Electrophysiol. 2002 Jun;25(6):967-83. doi: 10.1046/j.1460-9592.2002.00967.x.
5
Automatic mode switching of implantable pacemakers: II. Clinical performance of current algorithms and their programming.植入式起搏器的自动模式切换:II. 当前算法及其编程的临床性能
Pacing Clin Electrophysiol. 2002 Jul;25(7):1094-113. doi: 10.1046/j.1460-9592.2002.01094.x.
6
Improved efficacy of mode switching during atrial fibrillation using automatic atrial sensitivity adjustment.使用自动心房灵敏度调整改善心房颤动期间模式切换的疗效。
Pacing Clin Electrophysiol. 1999 Jan;22(1 Pt 1):17-25. doi: 10.1111/j.1540-8159.1999.tb00295.x.
7
Initial experience with a new algorithm for automatic mode switching from DDDR to DDIR mode.一种用于从DDDR模式自动切换到DDIR模式的新算法的初步经验。
Pacing Clin Electrophysiol. 1994 Nov;17(11 Pt 2):1908-12. doi: 10.1111/j.1540-8159.1994.tb03772.x.
8
Influence of autothreshold sensing and sinus rate on mode switching algorithm behavior.自动阈值感知和窦性心率对模式转换算法行为的影响。
Pacing Clin Electrophysiol. 2000 Oct;23(10 Pt 1):1473-8. doi: 10.1046/j.1460-9592.2000.01473.x.
9
Dynamic and dual-site atrial pacing in the prevention of atrial fibrillation: The STimolazione Atrial DInamica Multisito (STADIM) Study.动态双部位心房起搏预防心房颤动:动态多部位心房刺激(STADIM)研究
Pacing Clin Electrophysiol. 2007 Jan;30 Suppl 1:S71-4. doi: 10.1111/j.1540-8159.2007.00609.x.
10
[Mode-switching algorithms: programming and usefulness].[模式切换算法:编程与实用性]
Herz. 2001 Feb;26(1):2-17. doi: 10.1007/pl00002003.

引用本文的文献

1
Enhanced detection of atrial tachyarrhythmias with pacing devices by using more accurate atrial sensing.利用更精确的心房感知增强起搏设备检测房性心动过速。
J Interv Card Electrophysiol. 2022 Apr;63(3):601-609. doi: 10.1007/s10840-021-01066-z. Epub 2021 Oct 1.
2
Shortening of the sensed AV delay of a dual chamber pacemaker during normal sinus rhythm.双腔起搏器在正常窦性心律时感知到的房室延迟缩短。
J Interv Card Electrophysiol. 2003 Aug;9(1):29-34. doi: 10.1023/a:1025316419920.