Boellaard R, van Lingen A, van Balen S C, Hoving B G, Lammertsma A A
Clinical PET Centre, University Hospital Vrije Universiteit, Amsterdam, The Netherlands.
Eur J Nucl Med. 2001 Jan;28(1):81-9. doi: 10.1007/s002590000405.
The first performance tests of a new fully programmable blood sampling device for monitoring blood radioactivity during positron emission tomography (PET) are described. Blood is withdrawn through 1-mm internal diameter tubing using an infusion pump which can be operated at rates varying from 0 to 600 ml/h. Activity in blood is measured by a 6-cm-thick bismuth germanate crystal connected to a photomultiplier tube and multichannel analyser (MCA) which are positioned within 6 cm lead shielding. Positioning of the tubing is an exact and simple procedure. The minimal readout time of the MCA is 1 s. Two independent energy windows can be set. Operation of the pump and MCA is fully controlled by a PC, i.e. sampling time, interval time and pump rate can be varied at any time during the PET scan by user-defined scripts. A number of characteristics of the new system were studied, such as sensitivity, dead time, linearity, effect of background radiation and pump rate as a function of input pressure. In addition, dispersion was measured as a function of pump rate. Finally, first clinical results were compared with manual samples. The sensitivity equalled 0.7 and 0.2 cps/Bq for 511- and 1022-keV 30% energy windows, respectively, and the system dead time was 500 ns. The system remained linear within 2% with activity concentrations up to 2.5 MBq/cc. Short-term reproducibility was better than 3% for a 1-h period. Long-term reproducibility was about 5% (ISD), which was mainly caused by variation in the diameter of the tubing. If the device was positioned in such a way that maximum shielding was directed towards the patient, the effects of background radiation from the patient on the measured activity concentration for clinically relevant conditions was minimal (<3%). Pump rate varied with input pressure, but remained constant for a given pressure. Dispersion constants smaller than 0.14 s(-1) were observed for pump rates higher than 300 ml/h, indicating that the system dispersion is small. Clinical data showed an excellent agreement to within 3% (ISD) between the results obtained with the new system and manual samples. With the continuous blood sampler radioactivity in blood can be measured accurately during the entire course of the PET scan. Furthermore, the system is fully programmable allowing adjustment of all parameters during a single PET scan.
本文描述了一种新型全可编程血液采样装置的首次性能测试,该装置用于在正电子发射断层扫描(PET)期间监测血液放射性。使用输液泵通过内径为1毫米的管道抽取血液,输液泵的运行速率可在0至600毫升/小时之间变化。血液中的放射性通过一块6厘米厚的锗酸铋晶体进行测量,该晶体与一个光电倍增管和多道分析仪(MCA)相连,它们位于6厘米厚的铅屏蔽内。管道的定位是一个精确且简单的过程。MCA的最小读出时间为1秒。可以设置两个独立的能量窗口。泵和MCA的操作完全由一台个人计算机控制,即采样时间、间隔时间和泵速率可以在PET扫描期间通过用户定义的脚本随时更改。研究了新系统的一些特性,如灵敏度、死时间、线性度、本底辐射的影响以及泵速率随输入压力的变化。此外,还测量了分散度随泵速率的变化。最后,将首次临床结果与手动采集的样本进行了比较。对于511千电子伏和1022千电子伏的30%能量窗口,灵敏度分别为0.7和0.2计数每秒/贝克勒尔,系统死时间为500纳秒。在活性浓度高达2.5兆贝克勒尔/立方厘米的情况下,系统的线性度在2%以内。在1小时内,短期重现性优于3%。长期重现性约为5%(标准差),这主要是由管道直径的变化引起的。如果将该装置放置成使最大屏蔽朝向患者,在临床相关条件下,患者的本底辐射对测量的活性浓度的影响最小(<3%)。泵速率随输入压力而变化,但在给定压力下保持恒定。对于高于300毫升/小时的泵速率,观察到分散常数小于0.14秒-1,这表明系统分散度较小。临床数据显示,新系统获得的结果与手动样本之间的一致性极佳,在3%(标准差)以内。使用连续血液采样器,可以在PET扫描的整个过程中准确测量血液中的放射性。此外,该系统完全可编程,允许在单次PET扫描期间调整所有参数。