Kondera-Anasz Z, Michalski A, Gil D, Gil B, Starzewski J, Gonciarz Z
Department of Immunology and Serology, Silesian Medical University, Katowice, Poland.
Med Sci Monit. 2000 May-Jun;6(3):616-7.
The aim of this study was to establish the accuracy of determination of fibrinolytic proteins in human bile using commercial kits.
Gall bladder bile was obtained from patient undergoing laparoscopic cholecystectomy. The concentrations of tissue plasminogen activator (tPA), urokinase (uPA) and plasminogen activator inhibitor-1 (PAI-1) and inhibitor-2 (PAI-2) in bile were measured by enzyme linked immunoassay. Afterwards we conducted so called 'recovery test'.
Recovery percents of t-PA, u-PA, PAI-1, PAI 2 were very high i.e. 91.94, 94.07, 93.95, 91.2. We concluded that human fibrinolytic proteins in bile could be detected very accurately using commercial kits.