J Chromatogr A. 2000 Dec 22;904(1):113-7. doi: 10.1016/s0021-9673(00)00909-2.
A high-performance liquid chromatographic method with a dynamically modified amino column and evaporative light-scattering detector (ELSD) was established for the direct analysis of the carbohydrates in some drinks. A separation column (Zorbax Rx-SIL, 250 mm x 4.6 mm I.D., 5 microm, Hewlett-Packard, USA) which was modified by ethylenediamine and a guard column (Zorbax Rx-SIL, 12.5 mm x 4.6 mm I.D., 5 microm) were used. The mobile phase was a mixture of water-acetonitrile (1:2.6, v/v) containing 0.03% (v/v) ethylenediamine. Regression equations revealed linear relationship (correlation coefficients=0.996-0.999) between the mass of carbohydrates injected and the carbohydrates peak areas detected by ELSD. The detection limits of ELSD (S/N=3) were between 0.2 and 1.2 microg for different carbohydrates. This method is simple and sensitive.
建立了一种采用动态改性氨基柱和蒸发光散射检测器(ELSD)的高效液相色谱法,用于直接分析某些饮料中的碳水化合物。使用了经乙二胺改性的分离柱(Zorbax Rx-SIL,250 mm×4.6 mm内径,5微米,美国惠普公司)和保护柱(Zorbax Rx-SIL,12.5 mm×4.6 mm内径,5微米)。流动相为含有0.03%(v/v)乙二胺的水-乙腈混合物(1:2.6,v/v)。回归方程表明,进样的碳水化合物质量与ELSD检测到的碳水化合物峰面积之间呈线性关系(相关系数=0.996 - 0.999)。对于不同的碳水化合物,ELSD的检测限(S/N = 3)在0.2至1.2微克之间。该方法简单且灵敏。