Pei J, Li X
Department of Chemistry, The Hong Kong University of Science and Technology, Clear Water Bay, Kowloon.
Fresenius J Anal Chem. 2000 Aug;367(8):707-13. doi: 10.1007/s002160000447.
An amperometric sensor for amino acids based on the immobilization of amino acid oxidase on the surface of a CuPtCl(6)/GC is described. The amperometric current is due to the oxidation of H2O2 liberated during the enzyme reaction on the surface of the enzyme electrode. The electrode response characteristics as well as kinetic parameters have been evaluated. The enzyme electrode was characterized as an electrochemical biosensor, which was used as detector in high performance liquid chromatography (HPLC) for the determination of a mixture of amino acids with satisfactory results.
描述了一种基于将氨基酸氧化酶固定在CuPtCl(6)/GC表面的氨基酸安培传感器。安培电流是由于酶反应过程中在酶电极表面释放的H2O2的氧化。评估了电极的响应特性以及动力学参数。该酶电极被表征为一种电化学生物传感器,用作高效液相色谱(HPLC)中的检测器来测定氨基酸混合物,结果令人满意。