Glushchuk S F, Pekker Ia S
Med Tekh. 2001 Jan-Feb(1):12-5.
The paper presents data obtained from processing experimental findings of the electric conductivity of the actual media of the human gastrointestinal tract (GIT) as pH and temperature functions. Based on these data and experimentally derived values of electrode capacities of self-contained GIT electrostimulators in the real digestive tract, the authors calculated the values of stimulating pulse discharges. It was concluded that the discharge accumulated in the electrode containers is quite sufficient for stimulation of intestinal smooth muscle cells by using the purely capacity component of a stimulating pulse, It was shown that in this case the stimulating pulse was to be 0.8-1.2 msec.
本文介绍了通过处理人体胃肠道(GIT)实际介质的电导率随pH值和温度变化的实验结果所获得的数据。基于这些数据以及在实际消化道中自含式GIT电刺激器电极电容的实验推导值,作者计算了刺激脉冲放电的值。得出的结论是,通过使用刺激脉冲的纯电容分量,电极容器中积累的放电足以刺激肠道平滑肌细胞。结果表明,在这种情况下,刺激脉冲应为0.8 - 1.2毫秒。