Suppr超能文献

Quantitative analysis of regional wall motion and thickening by quantitative gated SPECT: comparison with visual analysis.

作者信息

Konno M, Morita K, Adachi I, Ito Y, Kohya T, Kitabatake A, Tsukamoto E, Tamaki N

机构信息

Department of Nuclear Medicine, Hokkaido University School of Medicine, Sapporo, Japan.

出版信息

Clin Nucl Med. 2001 Mar;26(3):202-7. doi: 10.1097/00003072-200103000-00004.

Abstract

PURPOSE

Electrocardiograph-gated myocardial SPECT permits a quantitative assessment of global and regional functions by quantitative gated SPECT software. To validate quantitative indexes of wall motion and wall thickening, the authors compared these indexes with visual scores.

MATERIALS AND METHODS

Gated myocardial SPECT was performed 60 minutes after the administration of Tc-99m sestamibi at rest in 42 patients with coronary artery disease. Regional wall motion (measured in millimeters and wall thickening (expressed as a percentage) were calculated by quantitative gated SPECT software in nine left ventricular myocardial segments and the results were compared with the five-point visual score interpretations of cinematic display.

RESULTS

A high correlation of wall motion was observed between the quantitative and visual analyses (r = 0.810; P < 0.001). In addition, a high but somewhat less significant correlation of wall thickening was observed between the quantitative and visual analyses (r = 0.606; P < 0.001).

CONCLUSIONS

In conclusion, regional wall motion and wall thickening can be evaluated quantitatively by electrocardiograph-gated myocardial SPECT and quantitative gated SPECT software. This will be useful for functional assessments made with various interventions.