Optical coherence tomography for the noninvasive evaluation of the cornea.

作者信息

Hirano K, Ito Y, Suzuki T, Kojima T, Kachi S, Miyake Y

机构信息

Department of Ophthalmology, Nagoya University School of Medicine, Aichi, Japan.

出版信息

Cornea. 2001 Apr;20(3):281-9. doi: 10.1097/00003226-200104000-00009.

Abstract

PURPOSE

To determine the optical coherence tomographic (OCT) characteristics of normal corneas and to characterize the OCT images of abnormal corneal lesions.

METHODS

Eleven eyes from 10 patients were examined at the Cornea Service of the Nagoya University Hospital: 4 had corneal pathologies, 4 underwent keratoplasty, and 2 were normal controls; 1 enucleated eye was also examined. OCT (OCT 2000 Zeiss-Humphrey) was used to study the normal cornea and various corneal abnormalities. We compared the OCT images to the observations made by slit-lamp biomicroscopy.

RESULTS

Fluid spaces were detected as black images. A highly reflective reflex was observed at the interface of different tissues, and intensive backscattering (reflex) was seen when the incident ray hit the laminated layers vertically. Corneal opacities were not clearly imaged when they were diffuse and mild, or when they were arranged axially in a small area, as was the scar of the graft-host junction. It was possible to obtain images from the region of the cornea that was not clearly visible by slit-lamp examination because of a corneal opacity.

CONCLUSION

OCT is a noncontact and noninvasive technique that can be performed safely on diseased corneas. OCT can provide objective documentation of corneal disorders that cannot be obtained by slit-lamp examination. The use of OCT in conjunction with other conventional instruments should provide a more complete image of the cornea.

摘要

目的

确定正常角膜的光学相干断层扫描(OCT)特征,并描述异常角膜病变的OCT图像特征。

方法

对名古屋大学医院角膜科的10例患者的11只眼睛进行了检查:4例有角膜病变,4例行角膜移植术,2例为正常对照;还检查了1只摘除的眼球。使用OCT(OCT 2000蔡司-汉弗莱)研究正常角膜和各种角膜异常情况。我们将OCT图像与裂隙灯生物显微镜检查的结果进行了比较。

结果

液性间隙表现为黑色图像。在不同组织的界面处观察到高反射性反射,当入射光线垂直照射到分层时可见强烈的后向散射(反射)。角膜混浊在弥漫性且轻度时,或在小区域内呈轴向排列时,如移植-宿主交界处的瘢痕,成像不清晰。由于角膜混浊,通过裂隙灯检查无法清晰看到的角膜区域也能够获得图像。

结论

OCT是一种非接触、非侵入性技术,可安全地用于患病角膜。OCT能够提供裂隙灯检查无法获得的角膜疾病的客观记录。将OCT与其他传统仪器结合使用应能提供更完整的角膜图像。