Nakonezny P A, Kowalewski R B, Ernst J M, Hawkley L C, Lozano D L, Litvack D A, Berntson G G, Sollers J J, Kizakevich P, Cacioppo J T, Lovallo W R
Veterans Affairs Medical Center and University of Oklahoma Health Sciences Center, Oklahoma City 73104, USA.
Psychophysiology. 2001 May;38(3):465-73.
The validity and reliability of a new ambulatory impedance cardiograph (AZCG) was tested against the Minnesota Impedance Cardiograph (ZCG) during rest, orthostasis, and mental stress. Impedance cardiography allows noninvasive assessment of stroke volume, cardiac output, and systolic time intervals. A reliable ambulatory device would allow studies outside the lab. The devices were compared at two sites in healthy subjects. In both studies, the AZCG tracked changes across conditions closely with the ZCG (all Period x Device interactions were nonsignificant). Pearson rs, were .65 to .93, random intraclass correlation coefficients ranged from .80 to .98, indicating high degrees of shared measurement variance, and Cronbach's alpha indicated very good internal reliabilities (.91 to .99). Relative to the ZCG, the new AZCG appears to provide valid and reliable estimates of cardiac function at rest and during behavioral challenges in the lab.
在静息、直立位和精神应激状态下,对一种新型动态阻抗心动图仪(AZCG)与明尼苏达阻抗心动图仪(ZCG)的有效性和可靠性进行了测试。阻抗心动图可对每搏输出量、心输出量和收缩期时间间期进行无创评估。一种可靠的动态设备将使实验室外的研究成为可能。在两个地点对健康受试者使用这两种设备进行了比较。在两项研究中,AZCG与ZCG在不同条件下的变化密切相关(所有时期×设备的交互作用均无统计学意义)。皮尔逊相关系数rs为0.65至0.93,随机组内相关系数范围为0.80至0.98,表明测量方差的共享程度很高,克朗巴哈系数表明内部可靠性非常好(0.91至0.99)。相对于ZCG,新型AZCG似乎能在静息状态以及实验室行为挑战期间对心功能提供有效且可靠的评估。