在体外试验中测量骨水泥型髋关节假体的移位情况。

Measurement of the migration of a cemented hip prosthesis in an in vitro test.

作者信息

Maher S A, Prendergast P J, Lyons C G

机构信息

Bioengineering Group, Department of Mechanical Engineering, Trinity College, Parsons Building, 2, Dublin, Ireland.

出版信息

Clin Biomech (Bristol). 2001 May;16(4):307-14. doi: 10.1016/s0268-0033(00)00116-9.

Abstract

OBJECTIVE

To develop a method to measure the migration of a cemented hip prosthesis in an in vitro experimental test.

DESIGN

A device to measure prosthesis movement relative to bone was designed and fabricated. It was tested using a Lubinus prosthesis (W. Link, Germany) implanted in a composite femur.

BACKGROUND

Clinical studies using radiostereophotogrammetry have shown that those cemented hip prosthesis that migrate rapidly in the first two post-operative years are the ones that require early revision. If migration be used as a basis for a pre-clinical test, then it should be possible to screen-out inferior designs before implantation in animal or clinical trials.

METHODS

The micromotion measurement device consisted of a 'target' of three spheres arranged in a cruciform structure. Six linear variable displacement transducers were aligned with the spheres so that motion of the prosthesis relative to the bone could be measured.

RESULTS

The displacement and rotation of the prosthesis relative to the composite femur was recorded for two million cycles. Relative rapid initial migration was found, followed by a period of steady-state migration. Distal migration (called 'subsidence' in this paper) of up to 0.1 mm was measured; however the variability in absolute prosthesis migration was very high despite efforts to ensure that all extraneous factors were minimised. In the majority of cases, the prostheses migrated medially, distally and anteriorly. The absolute subsidence, and its variability, were similar to that recorded clinically.

CONCLUSIONS

A method has been designed and tested which measures prosthesis migration in an experimental test. It provides a basis for a pre-clinical testing standard. Relevance. Hip prostheses need to be tested experimentally before implantation. However, no reliable test exists for such experimental tests. Rapid migration of a cemented prosthesis relative to bone has been shown in vivo to correlate with early failure, and in this paper a method to make such migration measurements in vitro is described and tested.

摘要

目的

开发一种在体外实验测试中测量骨水泥型髋关节假体移位的方法。

设计

设计并制造了一种测量假体相对于骨骼运动的装置。使用植入复合股骨的Lubinus假体(德国W. Link公司)对其进行测试。

背景

使用放射立体摄影测量法的临床研究表明,那些在术后头两年内快速移位的骨水泥型髋关节假体是需要早期翻修的假体。如果将移位用作临床前测试的基础,那么应该能够在植入动物或临床试验之前筛选出劣质设计。

方法

微动测量装置由一个呈十字形结构排列的三个球体的“靶标”组成。六个线性可变位移传感器与球体对齐,以便能够测量假体相对于骨骼的运动。

结果

记录了假体相对于复合股骨两百万次循环的位移和旋转。发现了相对快速的初始移位,随后是一段稳态移位期。测量到的远端移位(本文中称为“下沉”)高达0.1毫米;然而,尽管努力将所有外部因素降至最低,但假体绝对移位的变异性仍然非常高。在大多数情况下,假体向内、向远端和向前移位。绝对下沉及其变异性与临床记录的相似。

结论

设计并测试了一种在实验测试中测量假体移位的方法。它为临床前测试标准提供了基础。相关性。髋关节假体在植入前需要进行实验测试。然而,目前尚无用于此类实验测试的可靠方法。体内研究表明,骨水泥型假体相对于骨骼的快速移位与早期失败相关,本文描述并测试了一种在体外进行这种移位测量的方法。